Full digital ultrasonic diagnostic instruments (Vet) User’s Manual V2.00
20
5.
Press
key, the “
”mark appears on the screen. Hold down
or
key to change the
minor axis of the ellipse so as to satisfy the test area. The measured values are displayed at the
built-in characters “C: 00000mm, A: 00000mm
2
, V: 00000 cm
3
” on the right part of the screen
automatically;
6.
Press
key again to quit the minor axis status; Press
key or repeatedly click left mouse to
exchange the starting point and end point;
7.
Press
key or click right mouse to finish the first measurement;
8.
Repeat the steps from 3 to 7 to complete the multi-group data measurement;
9.
Continuously press
key twice to quit the measurement status;
10.
Press direction keys
to choose “Clear” in display area and press
key or synchronously click
the left and right key of mouse to clear all marks and data;
11.
Press
key or click the middle mouse to unfreeze, clear all marks and data and quit the
measurement status.
7.4 Slope/Heart rate/Cycle Measurement
The method to measure slope/hear rate/cycle is identical with distance measurement.
Note: In B/M mode, if both starting point and end point of the measurement mark fall into the
B-mode area, the value of the “+: ”refers to distance; if starting point and end point of the
measurement mark fall into the M-mode area, the value of the “+: ”refers to depth; if the
starting point and end point are in separate areas, the “+: ”will display “----”sign or
invalid value.
+
+
+
+
: denotes depth measured in mm (millimeter)
EF: denotes slope coefficient measured in mm/s (millimeter per second)
HR: denotes heart rate measured in times/minute (times per minute)
T: denotes cycle measured in ms (millisecond)
Attention:
The accuracy of software measurement: distance measurement ≤ 1mm;
area measurement ≤ 1mm
2
; volume measurement ≤ 1cm
3
; heart rate measurement ≤
1bpm; time measurement ≤ 1ms. Due to differences in images obtained by each user in
different times, the actual object for the accuracy of the measurement may be greater
than the above-mentioned values.