Full digital ultrasonic diagnostic instruments (Vet) User’s Manual V2.00
7
2. Remove the battery
According to the “OPEN” direction of the arrow, use the “Battery lock coin” to toggle “Battery cover
knob”, removed “Battery cover”, take out the battery.
Battery
Battery cover
Battery cover knob
5.4 Video recorder installation
1. Turn off the system, connect the equipotential terminal ( ) of the video recorder to the earthing;
2. Connect one end of the video connection line to the video recorder and the other end to the video
output interface on the left of main unit;
3. Insert one end of power plug (jack) of the video recorder to the power input socket of the video
recorder, the other end to the power supply socket.
5.5 Printer installation
1. Shutdown the system, insert printer data line into USB interface of printer, the other end to USB
interface of the main unit.
2. Insert one end of printer power plug (jack) into power input socket of the printer, and the other end to
the power supply socket.
Explanation:
The software version supporting the printer models
are tentatively scheduled for: HP Color
LaserJet CP1515n, HP Color LaserJet CP1525n. Supported printer models are subject to
change without notice!
Print encountered no paper, please restart the printer.
5.6 Connect to the mouse
Connect the mouse to USB interface in the side of the main unit.
5.7 Connect to power
1. Connect to the power adapter
Insert the output plug of adapter into DC power input interface.
2. Connect to the main power supply
Insert the power plug (jack) furnished with the machine into power input socket of the power adapter,
the other end to the mains socket-outlet. The instrument uses three-core power supply. It connects
with the protective earth line when power plug inserts into its socket.
5.8 Ultrasonic probe check before and after operation
Before and after ultrasonic diagnosis to check if there are any exceptionally on the surface of the probe
or cable jacket, such as peeling, cracks, bulge, or if the acoustic lens is reliable, disinfected or cleaned.
Warning:
1. Adapter has no switch. The isolation of the system with the MAINS used to unplug the
adapter as the intended isolation means.
2. The device should be used only with power adapter provided by Kaixin Company.
3. To avoid damaging power adapter or harming people by unexpected fallen, make sure the
power adapter is placed on the leveled desk.
4. The operator must not touch signal input/ signal output and patient simultaneously.
Attention:
USB port maximum loading current: 0.3A.