Full digital ultrasonic diagnostic instruments (Vet)                                                                                                    User’s Manual V2.00 

 

12 

6.5.4 Far Field Gain Adjustment 

In real-time mode, press

key to enter main-menu and press direction keys

to move cursor to 

“Far” in the display area. Press direction key

to increase far field gain and direction key

to reduce far 

field gain so as to control the gain in far field region.   

6.5.5 Dynamic Range Adjustment 

In real-time mode, press

key to enter main-menu and press direction keys

to move cursor to 

“Dyn”  in  the  display  area.  Press  direction  key

to  increase  the  value  of  dynamic  range  and  direction 

key

to decrease the value of dynamic range so as to control the dynamic range of the entire image. 

6.5.6 Frequency Adjustment (Frequency

 

conversion) 

In real-time mode, press

key to enter main-menu and press direction keys

to move cursor to 

“Freq.” in the display area. Press direction keys

to achieve frequency conversion. 

6.5.7 Focus Adjustment and Control 

In real-time mode and in B, B/B or 4B mode, press

key to enter main-menu and move cursor to “F. 

num”  and  then  press  direction  keys

to  choose  from  the  four  focus  modes  of  stage  1(Full  process 

dynamic focus), 2, 3 and 4.   

In  real-time  mode  and  in  B,  B/B,  4B,  B/M  or  M  mode,  press

key  to  quit  menu  mode,  press 

direction keys

again to move the focus up and down. 

Note: In B/M or M mode, only allowed choosing single focus mode. 

6.5.8 Frame Correlation Adjustment 

In real-time mode, press

key to enter main-menu and press direction keys

to move cursor to 

“Frame Avg” in the display area. Press direction keys

to achieve four levels of frame correlation. 

6.5.9 Image Post-process Adjustment 

In real-time mode, press

key to enter main-menu and press direction keys

to move cursor to 

“IP” in the display area and then press direction keys

to obtain corrected image. System default is 2.   

6.5.10 Edge Enhancement Adjustment 

In real-time mode, press

key to enter main-menu and press direction keys

to move cursor to 

“Edge”  in  the  display  area  and  then  press  direction  keys

to  gain  four  levels  of  sharpened  images. 

System default is 0.   

6.6 Image Control 
6.6.1 Magnification Selection 

In  real-time  mode,  press

key  to  enter  main-menu  and  move  cursor  to  “Zoom”  and  then  press 

direction keys

to choose eight kinds of magnification. 

6.6.2 Depth Range Selection and Depth Enhancement 

In real-time mode, press

key to enter main-menu and press direction keys

to move cursor to 

“Depth” in the display area and then press direction keys

to select depth, press

key to quit depth 

range selection. 

Note: There is no depth range selection function when main unit connecting the probe, which its 

nominal frequency greater than or equal to 6.5MHz.     

In real-time mode, press

key to enter main-menu and press direction keys

to move cursor to 

“Upgrade” in the display area and then press direction keys

, “

” appears on the screen. Press 

direction keys or operate mouse to move image up/down or left/right so as to observe the images of 

different depth and width. Press

key to quit depth heighten/pan status. 

Содержание RKU10

Страница 1: ...KAIXIN RKU10 Full Digital Ultrasonic Diagnostic Instruments Vet User s Manual XUZHOU KAIXIN ELECTRONIC INSTRUMENT CO LTD ...

Страница 2: ...out the express written permission of Kaixin Statement Kaixin has the final explanation right of this user s manual Kaixin was considered responsible for the safety reliability and performance in case of meeting all the following requirements 1 Assembly expansion readjustment improve and repair are all performed by professionals recognized by Kaixin 2 All replacement parts and accessories consumab...

Страница 3: ...r losses due to re installation alteration or repair the system by non Kaixin designated personnel 6 This product is intended to provide clinical diagnostic data for the doctor The doctor shall be responsible for the diagnostic process Kaixin shall not be liable for any problems arising out of the process 7 Be sure to back up important data to external storage media such as notebooks 8 Due to oper...

Страница 4: ...ger or unexpected use condition that may result in light injury or property loss or affecting the use if not avoided Indicates that make sure refer to relevant contents in this manual 2 Safety Symbols Symbols Meaning Symbols Meaning Type B applied part Keep dry Power supply indication Fragile DC Stacking limit by number Follow instructions for use Temperature limits Up Humidity limitation Marking ...

Страница 5: ...fter operation 7 5 9 Main unit check before and after operation 8 5 10 System reset 8 Chapter Six Functional Operation 9 6 1 Startup and Shutdown 9 6 2 Screen display reverse 9 6 3 System Functions Setting 9 6 3 1 Time Setting 9 6 3 2 TV Mode Setting 9 6 3 3 Energy Saving Setting 9 6 3 4 Characters Brightness Setting 9 6 3 5 Hospital Name Setting 9 6 3 6 Key Sound Setting 10 6 3 7 Compression Curv...

Страница 6: ...orage and recall 14 6 9 1 Save image 14 6 9 2 Open image 15 6 10 Single dump and mass dump 15 6 11 Delete the image 16 6 12 Image review 16 6 13 Print picture 17 6 14 Cine loop 17 6 15 Text Input 17 Chapter Seven General Measurement 18 7 1 Distance Measurement 18 7 2 Fat thickness Measurement 18 7 3 Circumference Area Volume Measurement 18 7 4 Slope Heart rate Cycle Measurement 20 Chapter Eight Ob...

Страница 7: ...ement function 7 Image freeze unfreeze 8 Depth range selection function 9 Depth heighten pan function 10 Color display function 11 Local zooming 12 Angle width change 13 Image filter function 14 Adjustment and display of near field far field total gain and dynamic range 15 Single point multi point focusing 16 Image storage 17 Vertical horizontal reverse and B W conversion 18 Case information image...

Страница 8: ...ens To convert electric signal to ultrasonic signal based on principle of converse piezoelectric effect The ultrasonic signal after entering the human body is reflected as echo wave and converted to electric signal again The acoustic lens is on the probe surface Supply ultrasonic coupling gel to the acoustic lens surface when performing ultrasonic diagnosis 2 Cable To connect the probe to the prob...

Страница 9: ...peration environment Ambient temperature 10 40 Relative humidity 30 75 without condensation Atmospheric pressure 700hPa 1060hPa Altitude 2000 m Overvoltage Overvoltage CategoryⅡ Pollution degree 2 4 3 Storage and transport Ambient temperature 20 55 Relative humidity 30 93 without condensation Atmospheric pressure 700hPa 1060hPa Danger Do not use this equipment where flammable gas such as anestheti...

Страница 10: ...may damage the instrument 7 The instrument is turned off completely only by disconnecting the power supply from the wall socket Warning 1 If breakers and fuse of the mains power socket are identical to those of this system and they are used to control the current for equipment like life support system the system shall not be connected to such power supply socket as it may cause breaker or fuse to ...

Страница 11: ...Hood 3 Fig Install the hood 2 The installed hood should be inserted the Hood buckle position aligned with the side of the instrument and then be inserted into the other side Note The force should be uniform 3 Take out the attached strap install it on the the hook of hood or put it on the the hole of strap of the instrument as shown in the figure below 4 Adjust the length of strap hanging on the ne...

Страница 12: ...remove the probe with figure s method and instructions Warning 1 The Kaixin ultrasonic probe shall be connected to the dedicated Kaixin ultrasonic system only Select proper probe model according to the relevant instructions of ultrasonic diagnostic system 2 Check the ultrasonic probe and connecting cable after diagnostic operation Use of defective probe may cause electric shock 3 Do not knock or b...

Страница 13: ...ls are subject to change without notice Print encountered no paper please restart the printer 5 6 Connect to the mouse Connect the mouse to USB interface in the side of the main unit 5 7 Connect to power 1 Connect to the power adapter Insert the output plug of adapter into DC power input interface 2 Connect to the main power supply Insert the power plug jack furnished with the machine into power i...

Страница 14: ...Check all the ports of the machine for possible damage or blockage 10 Clean the field and environment 5 9 2 Inspection after start up Check the following items after starting the machine 1 No abnormal voice strange smell and overheating appear 2 Check the machine to ensure a normal start up The power indication light is on and startup picture is shown on the screen Then the machine will be automat...

Страница 15: ... interface 3 Press direction keys to move symbol to point to TV mode 4 Press direction keys to achieve TV mode conversion between PAL and NTSC 5 Press key to quit setting interface 6 3 3 Energy Saving Setting 1 Press key to enter main menu and press direction keys to move cursor to Preset 2 Press direction keys to enter setting interface 3 Press direction keys to move symbol to point to Sleep 4 Pr...

Страница 16: ...ve 4 Press direction keys to select compression curve among 00 07 5 Press key to quit setting interface Note If use 3 5MHz probe it is recommended to choose 05 or 06 if use 5 0MHz probe recommend choose 03 or 04 if use 6 5MHz probe recommend choose 01 or 02 if use 7 5MHz probe recommend choose 00 or 01 6 3 8 Grid Setting 1 Press key to enter main menu and press direction keys to move cursor to Pre...

Страница 17: ...ultra scan mode Or right click mouse to achieve the change of scan mode 3 Change of M speed Press key to enter main menu and press direction keys to move cursor to M Speed and then press direction keys to select the eight kinds of scan speed Or left click mouse to switch M speeds 4 In real time mode press key to exit M mode 6 5 Image Quality Adjustment 6 5 1 Brightness and Contrast Adjustment In t...

Страница 18: ... time mode press key to enter main menu and press direction keys to move cursor to Frame Avg in the display area Press direction keys to achieve four levels of frame correlation 6 5 9 Image Post process Adjustment In real time mode press key to enter main menu and press direction keys to move cursor to IP in the display area and then press direction keys to obtain corrected image System default is...

Страница 19: ...menu and press direction keys to move cursor to Color in the display area and then press direction keys to achieve the conversion of eight kinds of colors including one kind of black and white 6 6 8 Scan Range Angle Width Change In real time B mode press key to enter main menu and press direction keys to move cursor to Angle and then press direction keys to choose angle width convex array for scan...

Страница 20: ...pt appears on lower right corner of the screen 3 Press direction keys to select the image code need to be saved in the LOCAL such as choose 003 4 Press key to appear Wait prompt After the prompt disappear the current image is saved in the frame for coded 003 The saved image code is preceded with asterisk 5 Press key to quit saving status and press key to return to real time status Storage to U dis...

Страница 21: ...r the number ID it would default to read out the images in the USER folder under the U disk 3 If has read out the images in the U disk also would like to re read the local image must press key to exit this reading status then press key to choose LOCAL and vice versa 6 10 Single dump and mass dump 6 10 1 Single dump From main unit to U disk 1 Read out the image in the main unit Img with highlight s...

Страница 22: ...o return to real time status Explanation 1 Mass dump which is dump images from main unit to DUMP folder in the U disk can also dump the images under the DUMP folder in the U disk to main unit 2 If U disk does not have DUMP folder or DUMP folder does not have images or the images within the DUMP folder are not named by digital 1 100 press key although the screen prompt Dumping the image number is n...

Страница 23: ... Press key unfreeze and quit playback status Note If the images appear abnormal that is without enough storage time and the images have not been stored full 6 15 Text Input Operation steps 1 Freeze image 2 Press key the cursor is located behind name 3 Press key again or right click mouse at the same time characters input menu will be shown at the bottom of the screen Caps 0 1 2 3 4 5 6 7 8 9 a b c...

Страница 24: ...all marks and data and quit the measurement status 7 2 Fat thickness Measurement 1 In B B B B M or M mode freeze the desired image the cursor is located in the Measure position of display area 2 Press key the measurement methods are showed in the lower left of the screen press direction keys to choose Distance press key again the cursor will show 3 Press key a dotted line of fat thickness measurem...

Страница 25: ...mouse to mark the first point continue to press direction keys or operate mouse move the mark to next desired position press key or click left mouse to mark the second point By analogy to mark all desired points Note The number of points must be within 8 to 32 4 When completed the marking of desired points press key or click right mouse it appears trace on the screen trace is automatically drawn a...

Страница 26: ...e clear all marks and data and quit the measurement status 7 4 Slope Heart rate Cycle Measurement The method to measure slope hear rate cycle is identical with distance measurement Note In B M mode if both starting point and end point of the measurement mark fall into the B mode area the value of the refers to distance if starting point and end point of the measurement mark fall into the M mode ar...

Страница 27: ...int of the measurement At the same time a lighted dotted line appears between the start and the end as the dashed locus of the measurement The measured value is automatically displayed at the built in mark mm on the right side of the screen G A and EDC value to be displayed in real time in the right area of the screen 6 Press key or repeatedly click left mouse to exchange the starting point and en...

Страница 28: ...o clear all marks and data 8 Press key or click the middle mouse to unfreeze clear all marks and data and quit the measurement status 8 4 Obstetric report In frozen status press key to enter main menu and press direction keys to move the cursor to the position Report Press direction keys to display obstetric report Press key to appear Wait prompt After the prompt disappear the obstetric report is ...

Страница 29: ... of regular disinfection Warning Turn off the instrument and pull out the power supply wire before cleaning every instrument of the system It may cause electric shock if clean the system underpoweris on Warning There is no any waterproof device in the system Do not splash any water or liquor into the system when cleaning or maintaining otherwise it will cause malfunction or electric shock Attentio...

Страница 30: ...dry cloth 2 In case of die hard blots clean with soft cloth dipped in moderate detergent and then air dry it 6 Clean the LCD screen Clean the liquid crystal display with dry soft flax or anti static LCD clean cloth Attention 1 Please carefully read the instructions provided by disinfectant provider about the disinfection liquid concentration and disinfection method as well as the description of th...

Страница 31: ...unit When the use time reduced significantly compared with normal conditions the battery should be promptly replaced 4 The excess high or low temperature will affect the charging and discharging performance and short the battery life and capacity Attention Battery is consumable the battery cycle life is based on the times of charge and discharge as unit When the use time reduced significantly comp...

Страница 32: ...e the battery which will lead to the leakage explosion or fire 8 Do not weld the battery terminal directly which will lead to the leakage explosion or fire 9 Do not disassemble the battery in any way which will lead to the leakage explosion or fire 10 Do not charge the battery near the heat source or extra hot environment which will lead to the leakage explosion or fire 11 Do not put the battery i...

Страница 33: ...safety checks should be performed at least every 12 months by a qualified person who has adequate training knowledge and practical experience to perform these tests Inspect the equipment and accessories for mechanical and functional damage Inspect the safety relevant labels for legibility Inspect the fuse to verify compliance with rated current and breaking characteristics Verify that the device f...

Страница 34: ... gas 3 Shockproof measures have been taken inside the packing case to allow for transport by air railway land and sea The goods shall not be exposed to poor weather conditions like rain and snow nor shall the goods be placed upside down bumped knocked or over stacked Chapter Eleven Safety Classification 1 Classified according to electric shock protection type Class I internally powered equipment 2...

Страница 35: ...in Mansion C 01 Economic Development Zone Xuzhou Jiangsu China Post Code 221004 Tel 86 516 87732932 87733758 Fax 86 516 87732932 87792848 Website http www kxele com E mail gm01 kxele com Information contained in this manual is subject to change without further notice ...

Отзывы: