Full digital ultrasonic diagnostic instruments (Vet) User’s Manual V2.00
19
3.
Press direction keys or operate mouse to move the “+” mark to desired position, press
key or
click left mouse to set the “+” mark position as the starting point of the measurement; or click right
mouse to set the “+” mark position as the starting point of the measurement;
4.
Press direction keys or operate mouse to move the“+” mark to the end point of the measurement.
At the same time, a locus appears in the direction of operation between the two measurement
marks. The measured circumference value is displayed automatically at the built-in mark “C:
00000mm” on the right part of the screen. Press
key or click right mouse to display at the
built-in mark “A: 00000mm
2
” the value of the measured area formed by measurement line
enclosure;
5.
Press
key, the measurement methods are showed in the lower left of the screen, continue to
choose “Trace” to measure;
6.
Repeat steps 3, 4 to complete the multi-group data measurement;
7.
Continuously press
key twice to quit the measurement status;
8.
Press direction keys
to choose “Clear” in display area and press
key or synchronously click
the left and right key of mouse to clear all marks and data;
9.
Press
key or click the middle mouse to unfreeze, clear all marks and data and quit the
measurement status.
Circumference/area measurement with multi-point method
1.
In B, B/B mode, freeze the desired image, the cursor is located in the “Measure” position of
display area;
2.
Press
key, the measurement methods are showed in the lower left of the screen, press direction
keys
to choose “Multi-point”, press
key again, the cursor will show “+”;
3.
Press direction keys or operate mouse to move the “+” mark to desired position, press
key or
click left mouse to mark the first point, continue to press direction keys or operate mouse, move
the “+” mark to next desired position, press
key or click left mouse to mark the second point; By
analogy, to mark all desired points;
Note: The number of points must be within 8 to 32.
4.
When completed the marking of desired points, press
key or click right mouse, it appears trace
on the screen (trace is automatically drawn along the order of marking point until close the
beginning point and end point), the measured circumference and area values are automatically
displayed at the built-in mark “C: 00000mm, A: 00000mm
2
” on the right part of the screen; If
continuously marks the thirty-two point, it directly appears trace on the screen, the measured
circumference and area values are automatically displayed at the built-in mark “C: 00000mm, A:
00000mm
2
” on the right part of the screen;
5.
Repeat steps 3, 4 to complete the multi-group data measurement;
6.
Continuously press
key twice to quit the measurement status;
7.
Press direction keys
to choose “Clear” in display area and press
key or synchronously click
the left and right key of mouse to clear all marks and data;
8.
Press
key or click the middle mouse to unfreeze, clear all marks and data and quit the
measurement status.
Circumference/area/volume measurement with ellipse method
1.
In B, B/B mode, freeze the desired image, the cursor is located in the “Measure” position of
display area;
2.
Press
key, the measurement methods are showed in the lower left of the screen, press direction
keys
to choose “Ellipse”, press
key again, the cursor will show “+”;
3.
Press direction keys or operate mouse to move the “+” mark to desired position, press
key or
click left mouse to set the “+” mark position as the starting point of the measurement;
4.
Press direction keys or operate mouse to move the“+” mark to the end point of the measurement, at
the same time the elliptic curve appears; or click middle mouse to appear elliptic curve;