Full digital ultrasonic diagnostic instruments (Vet) User’s Manual V2.00
10
Press direction keys or operate mouse to move cursor to point to Caps, and then press
key or
click left mouse to achieve capital and small letter conversion; If the cursor point to Shift,
press
key or click left mouse again to achieve the conversion between the letter and punctuation;
5.
Press direction keys to choose “numbers” or “characters” and press
key to confirm; Or operate
mouse to click “numbers” or “characters” to input;
6.
If need modify the content, press
key to quit the character input menu, press direction
keys
to delete input content; press
key again to retype;
7.
Press
key to save this setting and quit setting interface.
6.3.6 Key Sound Setting
1.
Press
key to enter main-menu and press direction keys
to move cursor to “Preset”;
2.
Press direction keys
to enter setting interface;
3.
Press direction keys
to move symbol “<” to point to “Key Sound”;
4.
Press direction keys
to select between “On” and “Off”;
5.
Press
key to quit setting interface.
6.3.7 Compression Curve Setting
1. Press
key to enter main-menu and press direction keys
to move cursor to “Preset”;
2. Press direction keys
to enter setting interface;
3. Press direction keys
to move symbol “<” to point to “Compression Curve”;
4. Press direction keys
to select compression curve among “00” –“07”;
5. Press
key to quit setting interface.
Note: If use 3.5MHz probe, it is recommended to choose 05 or 06; if use 5.0MHz probe,
recommend choose 03 or 04; if use 6.5MHz probe, recommend choose 01 or 02; if use
7.5MHz probe, recommend choose 00 or 01.
6.3.8 Grid Setting
1. Press
key to enter main-menu and press direction keys
to move cursor to “Preset”;
2. Press direction keys
to enter setting interface;
3. Press direction keys
to move symbol “<” to point to “Grid set”;
4. Press direction keys
to select “00”-“05”; “00” shows “no grid”, “01” shows “line grid”, “02”
shows “dot grid”, “03” shows 10mm-spacing dot-like scale, “04” shows 2mm-spacing dot-like scale,
“05” show 10mm-spacing and 2mm-spacing dot-like scales;
5. Press
key to quit setting interface.
Note: Grid is displayed on B mode after has been set.
6.3.9 Chinese-English Setting
1.
Press
key to enter main-menu and press direction keys
to move cursor to “Preset”;
2.
Press direction keys
to enter setting interface;
3.
Press
key to switch between Chinese and English;
4.
Press
key to quit setting interface.
6.4 Mode selection
In real-time mode, press
key repeatedly to achieve mode switching of B, B/B, 4B, B/M and M.
6.4.1 B Mode
B mode is a basic operation mode after startup and a single-framed B mode image is displayed.
6.4.2 B/B Mode
1.
In real-time mode, press
key to enter B/B mode.
2.
B/B image switch. Press
key to enter main-menu and press direction keys
to move the