Full digital ultrasonic diagnostic instruments (Vet) User’s Manual V2.00
18
Chapter Seven General Measurement
7.1 Distance Measurement
1.
In B, B/B or B/M mode, freeze the desired image, the cursor is located in the “Measure” position
of display area;
2.
Press
key, the measurement methods are showed in the lower left of the screen, press direction
keys
to choose “Distance”, press
key again, the cursor will show “+”;
(Note: In B/M
mode, measured method is distance measurement)
3.
Press direction keys or operate mouse to move the “+” mark to desired position, press
key or
click left mouse to set the “+” mark position as the starting point of the measurement;
4.
Press direction keys or operate mouse to move the“+” mark to the end point of the measurement. A
lighted dotted line appears between the starting point and the end point as the dashed locus of the
measurement. The measured value is automatically displayed at the built-in mark“+: ----mm” on
the right side of the screen;
5.
Press
key or click left mouse repeatedly to exchange the starting point and end point of the
measurement;
6.
Press
key or click right mouse to finish the first measurement;
7.
Repeat the steps 3~6 to complete the multi-group data measurement;
8.
Continuously press
key twice to quit the measurement status;
9.
Press direction keys
to choose “Clear” in display area and press
key or synchronously click
the left and right key of mouse to clear all marks and data;
10.
Press
key or click the middle mouse to unfreeze, clear all marks and data and quit the
measurement status.
7.2 Fat thickness Measurement
1.
In B, B/B, B/M or M mode, freeze the desired image, the cursor is located in the “Measure”
position of display area;
2.
Press
key, the measurement methods are showed in the lower left of the screen, press direction
keys
to choose “Distance”, press
key again, the cursor will show “+”;
3.
Press
key, a dotted line of fat thickness measurement appears in the middle of screen, press
direction keys
or operate the mouse to move the dotted line to desired position, press
direction keys
or operate the mouse to move the“+” mark to the end point of the
measurement, the measured value of fat thickness is automatically displayed at the built-in mark“+:
----mm” on the right side of the screen;
4.
Press
key or click right mouse to complete the multi-group fat thickness measurement;
5.
Press
key to return to the status of distance measurement;
6.
Continuously press
key twice to quit the measurement status;
7.
Press direction keys
to choose “Clear” in display area and press
key or synchronously click
the left and right key of mouse to clear all marks and data;
8.
Press
key or click the middle mouse to unfreeze, clear all marks and data and quit the
measurement status.
Note: Direction keys
cannot work in the fat thickness measurement when connecting
convex array probe.
7.3 Circumference/Area/Volume Measurement
Circumference/area measurement with trace method
1.
In B, B/B mode, freeze the desired image, the cursor is located in the “Measure” position of
display area;
2.
Press
key, the measurement methods are showed in the lower left of the screen, press direction
keys
to choose “Trace”, press
key again, the cursor will show “+”;