13 Error messages
99
13.2 Safety channel
Display
Origin
Cause/remedy
Terminal temperature
Internal
Terminal temperature outside the limits or probe faulty.
Reference A/D converter
Internal
Deviation during the reference measurement of the A/D converter /
Restart device/return device
Not calibrated
Internal
Channel not calibrated
Configuration
Internal
Configuration data outside the value range.
h
The message cannot be acknowledged until it is within the admissible range
again.
CRC Test Calibr.
Internal
Checksum error of the EEPROM calibration data.
Restart device/return device.
CRC Test Config.
Internal
Checksum error of the EEPROM configuration data.
Restart device/return device.
CPU Register
Internal
A CPU registry error has occurred.
Restart device/return device
RAM defective
Internal
A RAM error has occurred.
Restart device/return device.
ROM defective
Internal
A ROM error has occurred.
Restart device/return device.
Program sequence
Internal
A program sequence error has occurred.
Restart device/return device.
Содержание dTRANS T06 Ex
Страница 16: ...3 Mounting 16 3 Mounting 3 1 Dimensions...
Страница 30: ...5 Starting operation of the device 30 Here is a summary of the parameters from the figure above...
Страница 31: ...5 Starting operation of the device 31...
Страница 67: ...8 IECEx 67...
Страница 106: ...15 Certificates 106 15 Certificates...
Страница 111: ...15 Certificates 111 15 2 SIL and PL...
Страница 112: ...15 Certificates 112 15 3 ATEX...
Страница 113: ...15 Certificates 113...
Страница 114: ...15 Certificates 114...
Страница 121: ......
Страница 122: ......
Страница 123: ......