13 Error messages
100
Watchdog
Internal
A watchdog reset has occurred.
Restart device/return device.
Voltage supply
Internal
The voltage supply is insufficient.
Check voltage supply.
Frequency deviation
Internal
Error of the independent time base.
Restart device/return device.
EEPROM defective
Internal
Error during internal communication with the EEPROM.
Restart device/return device.
Stack
Internal
Error in the memory area reserved for the stack.
Restart device/return device.
A/D conversion
Internal
A program sequence error has occurred.
Restart device/return device.
Interrupt
Internal
Interrupt error
Restart device/return device.
Limit value
System
The configured limit value has been exceeded or undershot.
Check system/adapt limit value
Display
Origin
Cause/remedy