PAGE 56 of 89
6.3.5. 1 Axis Scan Mode
6.3.5.1 1 Axis Scan Setup Screen
1 Axis Scan Setup
Screen permits the user to setup the parameters of the
inspection.
NOTE:
Use the handheld controller to drive the crawler to the top dead centre of the
pipe before beginning the 1 Axis Scan Setup.
1
Diameter
The diameter of the pipe can be manually
entered. Alternatively, the diameter
is automatically calculated when the
circumference is entered.
2
Circumference
The circumference of the pipe can be
manually entered. Alternatively, the
circumference is automatically calculated
when the diameter is entered.
3
Shot count
The number of stops the crawler will
make around the pipe. Shot count will
automatically calculate the arc overlap.
4
Run
Opens
1 Axis Scan
Screen
where user can
start operation
(see “1 Axis Scan Screen” on
page 57)
.
5
Arc Overlap
Manually enter the amount of shot overlap
desired
(rounded to the nearest shot count)
.
Alternatively the overlap is automatically
calculated when the shot count is entered.
6
Speed
The speed when the crawler is travelling
between shot locations.
1 Axis Scan Setup
957.5 mm
Run
Circumference
10
Shot Count
304.8 mm
Diameter
209.1 mm
Arc Overlap
304.8 mm
Arc Length (Panel)
65 mm/s
Speed
Exit
1
2
3
4
8
6
5
7
Fig. 110 - 1 Axis scan mode