PAGE 4 of 89
Chapter 2
PRODUCT SPECIFICATIONS
2.1. Base SKOOT • X System
This section outlines the product specifications of the base system.
2.1.1. Intended Use
The
SKOOT • X
’s primary purpose is to perform inspections of assets such as
pipes, or small vessels by moving a detector and source around a surface. It is
intended for industrial use only.
2.1.1.1 Operating Limits
Category
Parameter
Specification
Inspection
Surface
Condition
Clean, free of excess rust, scale, debris
(i.e.
dirt, sand, etc.)
, ice, frost
Minimum OD
203 mm
(8.0 in)
Maximum OD
(Chain system)
1067 mm
(42 in)
Maximum surface temperature
50°C
(122°F)
Scanner
Maximum umbilical length
30 m
(100 ft)
Maximum source payload
(performance may vary with
surface conditions)
29 kg
(65 lbs)
2.1.1.2 Operating Environment
The SKOOT • X is for use in dry industrial environments having ambient
temperatures shown below. It is
NOT
intended for use in explosive
environments.
Category
Parameter
Specification
Environment
Minimum ambient
temperature
-20°C
(-4°F)
Maximum ambient
temperature
50°C
(122°F)