SECTION 18—FLOWMETER
Perfecto
2
™
70
Part No 1154245
SECTION 18—FLOWMETER
Replacing Flowmeter
DANGER
To prevent electrical shock, ALWAYS disconnect concentrator from electrical outlet
before servicing.
NOTE: For this procedure, refer to FIGURE 18.1 on page 70.
1. Unplug unit.
2. Remove cabinet. Refer to Removing Cabinet on page 23.
3. Note the location of the tubing for accurate replacement after new flowmeter is
installed.
4. Remove control panel. Refer to Removing Control Panel on page 60.
5. Remove palnuts that secure flowmeter to the control panel.
6. Remove flowmeter from front of control panel.
7. Install new flowmeter reversing STEPS 4-5.
8. Reinstall control panel. Refer to Removing Control Panel on page 60.
9. Run unit and check for leaks where 1/4-inch tubing engages flowmeter barbed fittings.
Refer to Leak Test on page 82.
10. Reinstall cabinet. Refer to Removing Cabinet on page 23.
FIGURE 18.1
Replacing Flowmeter
Palnuts
Control Panel
Flowmeter Barbed
Fittings (Back of
Flowmeter)
Flowmeter