SECTION 7—COMPRESSOR
Part No 1154245
47
Perfecto
2
™
Installing Compressor and Compressor Heads
NOTE: For this procedure, refer to FIGURE 7.8.
1. With the cylinder firmly seated on the compressor housing, replace both valve plate
o-rings and valve plates.
CAUTION!
Risk of Damage
–
Make sure that the valve plate o-rings are not twisted when seated onto the compressor
housing.
2. Replace both head seal o-rings and compressor head.
3. Insert head screws and tighten in a crisscross pattern to 4.0 N-M (35.4 inch-lbs).
WARNING!
Risk of Injury or Damage
–
Rotate the fan by hand prior to connecting the unit to a power source. Check for suction
at the air inlet by placing your finger over the port as you rotate the fan. You should feel
a slight suction with each rotation of the fan. If you don’t feel suction or the fan does not
turn freely, DO NOT connect the unit to a power source. Review the assembly procedure
for possible errors.
4. Install the compressor. Refer to Replacing Compressor Assembly on page 31.
5. Run the unit and check for leaks around compressor hoses and compressor tubes.
6. Take pressure readings to ensure proper system pressures.