SECTION 25—SIEVE BED PRESSURE CHECK
Part No 1154245
93
Perfecto
2
™
Checking Sieve Bed Pressure (Model IRC5PO2VAW
Only)
DANGER
To prevent electrical shock, ALWAYS disconnect concentrator from electrical outlet
before servicing.
Disassembly
NOTE: For this procedure, refer to FIGURE 25.3.
FIGURE 25.3
Disassembly
1. Turn power off ( ) and unplug unit.
2. Remove cabinet. Refer to Removing Cabinet on page 23.
3. Turn unit so the product tank faces you.
4. Remove tie wrap from the end of the 1/8” pressure sensor hose that connects to the
product tank.
5. Remove pressure sensor hose from product tank.
Installation
1. Insert "T" fitting into the end of the pressure sensor hose that was connected to the
product tank cap.
2. Attach a piece of 1/8” tubing to the opposite leg of the "T" fitting.
3. Connect the "T" fitting assembly to product tank where the pressure sensor hose was
connected.
4. Secure each end of the 1/8” tubing to the "T" fitting connection with hose clamps.
Check Valve
1/8” Pressure
Sensor Hose
attaches HERE
Product
Tank
/O