TYPICAL PRODUCT PARAMETERS
Part No 1154245
9
Perfecto
2
™
O
2
INDICATORS
Oxygen Output Temperature:
Less than A3°C (+ 8°F)
Cabinet:
Impact Resistant flame-retardant plastic cabinet that conforms to
UL 94-V0
No AP/APG:
Not suitable for use in the presence of a flammable anaesthetic
mixture.
Electrical:
No extension cords
Placement:
No closer than 30.5 cm (12 in) from any wall, furniture or
draperies to assure sufficient air flow. Avoid deep pile carpets and
heaters, radiators or hot air registers. Floor location only.
Tubing:
2 m (7 ft) cannula with a maximum 15 m (50 ft) of Crush-Proof
Tubing (DO NOT pinch)
Relative Humidity:
20 to 60%
Time of Operation:
Up to 24 hours per day
Recommended Storage and Shipping
Temperature:
-29°C to 65°C (-20°F to 150°F), at 15-95% relative humidity
Environment:
Smoke and soot-free. No confined spaces (Example: No closets).
Standards and Regulatory Listing:
IEC/EN61000-3-2
IEC/EN61000-3-3
IEC/EN 60601-1, A1, A2
IEC/EN 60601-1-2
ISO8359
MDD 93/42/EEC, Annex I and IX
CE marked model:
IRC5PO2AW/IRC5PO2VAW
Minimum Operating Time:
30 Minutes
LABEL
SYMBOL
O
2
PURITY
INDICATOR LIGHTS
(LED)
SYSTEM OKAY
O
2
over 85%
GREEN Indicator Light
O
2
Between 73% to 85% YELLOW Indicator light
A. YELLOW Solid
B. YELLOW Flashing Sensor
Failure
Call a qualified technician.
SYSTEM FAILURE
O
2
Below 73%
RED Indicator Light
Continuous Audible Alarm
Sieve-GARD Compressor Shut-
down.
O
2