SECTION 4—TROUBLESHOOTING
Perfecto
2
™
22
Part No 1154245
Unit Not Operating,
Internal Status Indicators:
RED: Off
GREEN: Off
Control Panel Indicators:
RED: Off
YELLOW: Off
GREEN: Off
Power Switch ON.
Continuous audible alarm.
Transformer assembly:
a. Assembly connector
disconnected.
b. Faulty transformer assembly.
a. Reattach connector.
b. Replace transformer assembly. Refer
to Replacing the Transformer on
page 68.
Unit Operating,
Internal Status Indicators:
RED: Three Flashes
GREEN: One Flash
Control Panel Indicators:
RED: Off
YELLOW: Flashing
GREEN: On
Internal repairs required.
Replace SensO
2
circuit board. Refer to
Replacing P.C. Board on page 65.
Potential Obstruction Alert
does not activate on flows
less than 0.5 L/min.
System leak.
Repair leak in product tank, regulator,
tubing, fittings, or flow meter.
Defective check valves.
Replace check valves. Refer to Replacing
Check Valves on page 53.
SYMPTOM
PROBABLE CAUSE
SOLUTION