SECTION 7—COMPRESSOR
Perfecto
2
™
40
Part No 1154245
Compressor
WARNING
To avoid personal injury or property damage, rotate the fan by hand prior to connecting
the unit to a power source. Check for suction at the air inlet by placing your finger over
the port as you rotate the fan. You should feel a slight suction with each rotation of the
fan. If you DO NOT feel a suction but you feel or hear a thump as you turn the fan, DO
NOT connect the unit to a power source. Review the assembly procedure for possible
error.
NOTE: For this procedure, refer to FIGURE 7.7 on page 41.
1. Reinstall compressor. Refer to Replacing Compressor Assembly on page 31.
2. Run unit and check for leaks around compressor hoses and compressor tubes. Refer
to Leak Test on page 82.
3. Take pressure readings to ensure proper system pressures. Refer to Checking Sieve
Bed Pressure (Model IRC5PO2AW Only) on page 91 or CHECKING SIEVE BED
PRESSURE (MODEL IRC5PO2VAW ONLY) on page 93.