TS25 MKII : TS25-BAS1-KIT : TS25-BAS1-CAN-240
Battery charger/analyser : Operators Manual
progress. If the
TEST
key is pressed this will briefly display the current test
mode.
4.2.5 STOPPING A CAPACITY TEST
At any time during the capacity test the process can be terminated by simply
pressing the
STOP
key. When the
STOP
key is pressed the process will be
terminated immediately and the display will show
END
and the
END
led will remain
illuminated.
The display will remain in this state until either
MUTE
of
STOP
is pressed.
4.2.6 CAPACITY TEST AUTO TERMINATE
If the capacity test has been programmed to auto terminate on time then the
process will automatically stop when either the battery voltage falls below the
voltage end point or when the capacity test time has elapsed.
If the capacity test has been programmed to auto terminate only when the
voltage falls below the voltage end point, then the process will stop when the
battery voltage falls below the voltage end point.
In both cases the capacity test process will terminate immediately, the audible
alarm will sound and the
ALARM
and
END
lights will illuminate and the display will
show the reason for terminating the process.
To clear the alarm the
MUTE
or
STOP
key must be pressed. Upon these being
pressed the display will show the actual capacity percent value.
Process termination codes:
E 1
The process time is complete.
E 2
Battery voltage has reached the voltage end point.