2 Safety information copper tests
8
ARGUS Copper Test Manual
Measuring accuracies copper tests:
Voltage:
DC voltage measurement (U=):
Measuring range:
Resolution:
Accuracy:
0 V - 9.99 V
0.01 V
±(0.5 % + 2 digits)
10 V - 200 V
0.1 V
±(0.5 % + 2 digits)
AC voltage measurement (sine, 50 Hz) (U~):
Measuring range:
Resolution:
Accuracy:
0 V - 9.99 V
0.01 V
±(2 % + 2 digits)
10 V - 200 V
0.1 V
±(1.5 % + 2 digits)
Frequency: 10 Hz to 200 Hz; 0.2 Hz; ±(1.5 % + 2 digits); sine
When measuring voltages, the input resistance depends on the measuring range:
- Measuring range 1: 0 V - 25 V; Input resistance: 1.694 MΩ
- Measuring range 2: 25 V - 200 V; Input resistance: 1.05 MΩ
As soon as the total of all voltages exceeds 25 V, the measurement is switched over.
Capacitive symmetry measurement (CSym):
Measuring range:
Resolution:
Accuracy of capacity diffe-
rence:
10 nF - 4 µF
0.01 nF
0.1% of the capacity to
ground
Capacitive measurement, measuring frequency 8 Hz (C):
Measuring range:
Resolution:
Accuracy:
0.01 nF - 9.99 nF
0.01 nF
±(4 % + 4 digits)
10 nF - 99.99 nF
0.01 nF
±(4 % + 4 digits)
100 nF - 999.9 nF
0.1 nF
±(3 % + 1 digit)
1 μF - 8 μF
1 nF
±(3 % + 1 digit)
* All specifications relate to a comparison measurement using film capacitors.