2 Safety information copper tests
10
ARGUS Copper Test Manual
Loop resistance measurement with 13 V, max. 15 mA (Iso.):
Measuring range:
Resolution:
Accuracy:
1
Ω
- 999.9
Ω
0.1
Ω
±(1 % + 3 digits)
1 k
Ω
- 9.999 k
Ω
1
Ω
±(1 % + 1 digit)
10 k
Ω
- 99.99 k
Ω
10
Ω
±(1 % + 1 digit)
100 k
Ω
- 999.9 k
Ω
100
Ω
±(1 % + 1 digit)
1 M
Ω
- 9.999 M
Ω
1 k
Ω
±(2 % + 1 digit)
10 M
Ω
- 40 M
Ω
10 k
Ω
±(5 % + 1 digit)
DC current measurement (I=):
Measuring range:
Resolution:
Accuracy:
0 mA - 150 mA
0.1 mA
±(2.5 % + 3 digits)
Unsymmetrie balance at 1 MHz (LCL):
Measuring range:
Resolution:
Accuracy:
0 dB - 55 dB
0.1 dB
± 1.5 dB
55.1 dB - 65 dB
0.1 dB
± 3 dB
The length of the measuring leads can significantly affect the measurement. Therefore,
always use the original accessories for measuring.
Near-end crosstalk at 1 MHz (NEXT):
Measuring range:
Resolution:
Accuracy:
0 dB - 65 dB
0.1 dB
± 1 dB
The length of the measuring leads can significantly affect the measurement. Therefore,
always use the original accessories for measuring.
Reference conditions (calibration):
- Temperature: +23 °C to ± 5 °C
- Humidity: 50 % ± 20 %, relative, non-condensing
- Frequency of measurement quantity: 50 Hz ± 5 Hz, sine