3 Copper tests
22
ARGUS Copper Test Manual
3.1.5 Isolation resistance measurement (Iso.)
Isolation resistance measurement can indicate damage to cable isolation, moisture
penetration or oxidised contact points on the line.
See the chapter Technical data page 9 for information on measuring ranges, resolutions
and accuracies of isolation resistance measurement.
Isolation resistance measurement settings
Before the measurement, the selected test parameters are displayed.
- Measuring voltage (in this example 105 V)
- Measuring current (in this example max. 1.05 mA)
- Polarity (in this example Automatic, to 1GΩ)
After selecting the "Menu" softkey, the following settings can be made:
- Polarity
- Measuring range
- Measuring voltage
- Susceptibility
- Remote kit command
- Remote kit
- Information bell
The configuration options can vary depending on the selected sockets.
A voltage check is performed before the measurement in which the voltage
limits given in the table (see page 7) may not be exceeded.