3 Copper tests
ARGUS Copper Test Manual
27
4. If an intact conductor is used as an auxilliary conductor instead of ground, RE=RH, RH is
then entirely on the order of magnitude of Ra and Rb. A complete twisted pair (Tip-Ring)
can also be used for this.
Calculation basis:
The following calculation basis is used for a 3-way measurement Rab, RaE, RbE, with R1,
R2, R3 as the total resistance of the respective line pairs.
Additionally, the absolute deviation of the two resistances Ra and Rb and the relative
deviation are determined.
Relative deviation (in %) = 2x
ΔR (absolute deviation in Ω) = absolute deviation (Ra - Rb)
See the chapter Technical data page 9 for information on measuring ranges, resolutions
and accuracies of resistance symmetry measurement.
A voltage check is performed before the measurement in which the
voltage limits given in the table (see page 7) may not be exceeded.
A loop must be created between Tip-Ring-Ground. This can be achieved
using the instrument kit function "Tip-Ring-Ground".