3 Copper tests
ARGUS Copper Test Manual
13
3.1.2 Voltage measurement (U= and U~)
Voltage measurement enables detection of e.g. connected accesses (such as POTS) and
supply voltages and external voltages, such as a connection to another line.
In AC voltage measurement, the frequency (10 - 200 Hz) is additionally displayed. This also
enables identification of call switching and coupled voltages as well as carrier or power fre-
quencies.
See the chapter Technical data page 8 for information on measuring ranges, resolutions
and accuracies of voltage measurement.
Selection of voltage measurement.
Voltage measurement starts automatically when the Copper tests are activated. In this
example, the
- Current voltage in volts
- Current frequency of AC voltage in hertz
- Minimum voltage in volts
- Maximum voltage in volts
are measured.
After selecting the "Menu" softkey, the following settings can be made:
- Input Resistance
- Remote Kit Command
- Remote Kit
- Information Bell
Note that unexpected or hazardous mixed voltages or hazardous
voltage peaks can occur.