Industrex M43ic / M43icN
page 39
01/2013 TK
Problem
8. Main drive and dryer
run continuously
9. Material wet when
exiting processor
10. Temperature problems
Temperature is shown
incorrect.
Possible cause
A) Main drive was started in
"manual mode"
B) Material always under
sensorbar. Material not
transported/pulled into the
processor
C) Sensor/s at the sensorbar
dirty
D) Main board defective
A) Dryer temperature too low
B) Transport speed to high
C) Unusable or wrong
Developer or Fixer
D) Dryer blows only cold air
The temperature probes
has to be positioned
according their coding.
correction
A) Check in the manual
programm if "STOP" is
shown stop the transport
with button.
Attention: if also an
automatic cycle is started by
the sensor bar this cycle will
end first.
B) Check the Input rubber
roller. Check the film
casette.
C) Clean the sensor/s
D) Call service/change main
board
A) increase the dryer
temperature (max. 60°C)
B) Lower the transport speed
C) Increase the Repl.rate or
change the chemicals
D) Fuse F1 of the dryer heater
defective or solid state relais
1. The temperature probeS are
colour coded
yellow
Developer
blue
Fixer
red
Dryer
Содержание M43ic
Страница 1: ...INSTRUCTION MANUAL 04 2011 TK NDT FILM PROCESSOR Industrex M43ic M43icN AT800 v2 8r09p and up ...
Страница 3: ...Input speed DEV time Cycle time KODAKIndustrexM43ic Industrex M43ic M43icN ...
Страница 49: ...page 43 01 2013 TK Industrex M43ic M43icN NOTES ...
Страница 50: ...page 44 01 2013 TK Industrex M43ic M43icN NOTES ...
Страница 51: ...page 45 01 2013 TK Industrex M43ic M43icN NOTES ...