5.6 ERRORS CODES
If an error occurs, the indication P1 (or M1) will alternate with Er. If this happens
Press the
2
button. This will stop the beeper and bring you to the error menu, so you can
check what kind of error is indicated. If more than one error occurred, press
8
to scan
the rest of them. Press
4
to jump back to main page.
When the filmprocessor switched on at the beginning of the working
hours ,it is expected to have low temperatures in the tanks. For this
reason, the
Er
indication will be present, but without alarm.
If, however, the temperature drops during normal work, the
alarm will be activated.
Error Messages:
Display
Reason
Tank1 too cold
A:Normal condition during heat-up period. The developer will be
heated until the preprogrammed-temperature is reached.
B:If an error message is displayed.
Call for qualified Service personnel
Tank1 too warm
Developer temperature has gone up more than 1°C above SET-
temperature.Call for qualified Service personnel
Tank2 too cold
A:Normal condition during heat-up period. The fixer will be
heated until the preprogrammed-temperature is reached.
B:If an error message is displayed:
Call for qualified Service personnel
Tank2 too warm
Fixer temperature has gone up more than 5°C above SET-
temperature.Call for qualified Service personnel
Dryer too warm
Actual temperature in the dryer is more than 5°C above SET-
temperature.Call for qualified Service personnel
Motor overload
Hardly running drive/transportsystem. The drive motor did not reach
it´s SET-speed Call for qualified Service personnel
Tank1 low level
Level Tank1 too low
Tank2 low level
Level Tank2 too low
Water overflow
Drain of watertank is blocked Call for qualified Service personnel
Cover openend
Top cover of the filmprocessor is open.
T1: no probe
Temperature probe in Tank1 is defect or lacks Call for qualified
Service personnel
T2: no probe
Temperature probe in Tank2 is defect or lacks Call for qualified
Service personnel
T3: no probe
Temperature probe in Dryer is defect or lacks Call for qualified
Service personnel
Water low level
Level Watertank is too low.
Can't fill water
After
20 minutes
, the watertank should be filled up with water, if
the sensor is not reached during this time, the meassage appears.
Call for qualified Service personnel
Wrong location
Only possible, if a FEEDER is installed - refer to the „Instruction
Manual for the Auto Film Feeder“.
Change filter
The filter medium has to replaced. Call for qualified Service
personnel
Errors: 1
Tank1 too cold
Содержание M43ic
Страница 1: ...INSTRUCTION MANUAL 04 2011 TK NDT FILM PROCESSOR Industrex M43ic M43icN AT800 v2 8r09p and up ...
Страница 3: ...Input speed DEV time Cycle time KODAKIndustrexM43ic Industrex M43ic M43icN ...
Страница 49: ...page 43 01 2013 TK Industrex M43ic M43icN NOTES ...
Страница 50: ...page 44 01 2013 TK Industrex M43ic M43icN NOTES ...
Страница 51: ...page 45 01 2013 TK Industrex M43ic M43icN NOTES ...