Industrex M43ic / M43icN
page 20
01/2013 TK
To set the parameters, move the cursor with
6
,
then change the value with
7
.
Once all digits have been set to the desired value, position the cursor under
Save and press
7
to store the values, or select Cancel to discard the changes.
If any of the values is set too high or too low, when trying to save the programme
you will get.
After 2 seconds the message will disappear and you will be taken back to re-programme the
values. A parameter that was too high will be automatically reset to the maximum
possible value. A parameter that was too low will be reset to the minimum
possible value.
This can be used if you want to program extreme values - for instance you want to use
the shortest developing time possible, but you don’t remember the value. In this case
just programme 000. After the “Out of range” message, the developing time will be reset to
the minimum. Just select Save once again.
To go back to work mode, press
4
.
.
Note: The Setup menu is just for service and factory setting purposes only.
The Setup-menu set filmprocessor-specific values. This values should not be changed by the costumer.
Although you can select these menus, scroll through the data and even change it, saving the
changed data is disabled. Only authorized service technician can re-programme these values.
Out of range
Содержание M43ic
Страница 1: ...INSTRUCTION MANUAL 04 2011 TK NDT FILM PROCESSOR Industrex M43ic M43icN AT800 v2 8r09p and up ...
Страница 3: ...Input speed DEV time Cycle time KODAKIndustrexM43ic Industrex M43ic M43icN ...
Страница 49: ...page 43 01 2013 TK Industrex M43ic M43icN NOTES ...
Страница 50: ...page 44 01 2013 TK Industrex M43ic M43icN NOTES ...
Страница 51: ...page 45 01 2013 TK Industrex M43ic M43icN NOTES ...