Industrex M43ic / M43icN
page 14
01/2013 TK
3.1 USING THE CHEMISTRIES
*)
Only use chemistry suitable for roller transport systems.
*)
Follow instructions of chemistry manufacturers.
FIXER BATH:
*)
Empty fixer tank by opening the fix drain tap.
*)
Remove the Fi
xer-
rack.
*)
Check fixer tank is free of alien material.
*)
Close fix drain tap.
*)
Fill fixer tank with ready-to-use-fixer solution to the red marker on the tank wall. Insert
the Fixer-rack very carefully and slowly, add hardener solution if advised by the chemistry
manufacturer.
DEVELOPER BATH:
*)
Empty developer tank by opening dev drain tap.
*)
Remove the Developer-rack.
*)
Check developer tank is free of alien material.
*)
Close dev drain tap.
*)
Fill developer tank with ready-to-use-developer solution to the red marker on the tank
wall. Insert the developer-rack very carefully and slowly. Replenishment tanks may be used
to mix the chemistry.Any remaining can be used for replenishment.
CAUTION:
Even the smallest quantity of fixer could contaminate the developer
solution. Therefore, always fill with fixer first. When removing the fixer
rack, always cover the developer tank. For removing the fixer rack
use rack carrier tray (optional accessory)
3. THE FIRST STEPS
WARNING: Separate the Film Processor from mains. To do so, switch the main
power switch of the Filmprocessor to „0“ position. Wear safety goggles,
protection gloves and clothing.
Содержание M43ic
Страница 1: ...INSTRUCTION MANUAL 04 2011 TK NDT FILM PROCESSOR Industrex M43ic M43icN AT800 v2 8r09p and up ...
Страница 3: ...Input speed DEV time Cycle time KODAKIndustrexM43ic Industrex M43ic M43icN ...
Страница 49: ...page 43 01 2013 TK Industrex M43ic M43icN NOTES ...
Страница 50: ...page 44 01 2013 TK Industrex M43ic M43icN NOTES ...
Страница 51: ...page 45 01 2013 TK Industrex M43ic M43icN NOTES ...