tM-TH8 User Manual, Rev: A1.1 7MH-018-A11
93
2.31 @AAGxTtt
Description:
Reads the Steinhart coefficient of a user-defined type.
Syntax:
@AAGxTtt[CHKSUM](CR)
@
Delimiter character
AA
Address of the module to be read (00 to FF)
G
Command to read the Steinhart coefficient
x
A: reads Steinhart coefficient A
B: reads Steinhart coefficient B
C: reads Steinhart coefficient C
Ttt
tt represents the type code of the user-defined
type to read the Steinhart coefficient.
Response:
Valid command:
!AA(Data)[CHKSUM](CR)
Invalid command:
?AA[CHKSUM](CR)
!
Delimiter character for a valid command
?
Delimiter character for an invalid command or
invalid type code
AA
Address of the responding module (00 to FF)
(Data)
Eight hexadecimal digits to represent the
Steinhart coefficient in IEEE-754 format. See
Section 1.11 for details.
There will be no response if the command syntax is
incorrect, there is a communication error, or there is no
module with the specified address.