tM-TH8 User Manual, Rev: A1.1 7MH-018-A11
62
tM-TH8
2.16 $AAP
Description:
Reads the communication protocol information.
Syntax:
$AAP[CHKSUM](CR)
$
Delimiter character
AA
Address of the module to be read (00 to FF)
P
Command to read the communication protocol
Response:
Valid Response:
!AASC[CHKSUM](CR)
Invalid Response: ?AA[CHKSUM](CR)
!
Delimiter character for a valid response
?
Delimiter character for an invalid response
AA
Address of the responding module (00 to FF)
S
The protocols supported by the module
0: only DCON protocol is supported
3: The DCON, Modbus RTU and Modbus
ASCII protocols are supported
C
Current protocol saved in EEPROM that will be
used at the next power on reset
0: the protocol set in EEPROM is DCON
1: the protocol set in EEPROM is Modbus RTU
3: the protocol set in EEPROM is Modbus
ASCII
There will be no response if the command syntax is
incorrect, there is a communication error, or there is no
module with the specified address.
Содержание tM-TH8
Страница 8: ...tM TH8 User Manual Rev A1 1 7MH 018 A11 8 1 2 Terminal Assignment...
Страница 10: ...tM TH8 User Manual Rev A1 1 7MH 018 A11 10 1 4 Block Diagrams 1 4 1 Block Diagram for the tM TH8...
Страница 11: ...tM TH8 User Manual Rev A1 1 7MH 018 A11 11 1 5 Dimensions...
Страница 13: ...tM TH8 User Manual Rev A1 1 7MH 018 A11 13 1 7 Quick Start Please refer to the Quick Start Guide for tM TH8...
Страница 67: ...tM TH8 User Manual Rev A1 1 7MH 018 A11 67 Related Topics Section 1 9 Calibration tM TH8...