tM-TH8 User Manual, Rev: A1.1 7MH-018-A11
87
2.28 @AAA3Ci
Description:
Reads the temperature offset of a channel.
Syntax:
@AAA3Ci [CHKSUM](CR)
@
Delimiter character
AA
Address of the module to be read (00 to FF)
A3 Command to read the temperature offset
Ci
i specify the input channel to be set (0 ~ 7)
.
Response:
Valid command:
!AA(Data)[CHKSUM](CR)
Invalid command:
?AA[CHKSUM](CR)
!
Delimiter character for a valid command
?
Delimiter character for an invalid command or
invalid type code
AA
Address of the responding module (00 to FF)
(Data)
Two hexadecimal digits to represent the
temperature offset. It is
01 for 0.1, 02 for 0.2, FF
for -0.1, FE for -0.2, etc.
There will be no response if the command syntax is
incorrect, there is a communication error, or there is no
module with the specified address.