tM-TH8 User Manual, Rev: A1.1 7MH-018-A11
71
Examples:
Command: ~01DC
Response: !01
Sets the temperature scale of module 01 to Celsius,
and returns a valid response.
Command: ~02DF
Response: !02
Sets the temperature scale of module 02 to
Fahrenheit, and returns a valid response.
Related Commands:
Section 2.19 ~AAD
Notes:
1.
The default temperature scale is Celsius.
2.
The commands affected are #AA, #AAN, and
@AARTTttR(Data).
tM-TH8
Содержание tM-TH8
Страница 8: ...tM TH8 User Manual Rev A1 1 7MH 018 A11 8 1 2 Terminal Assignment...
Страница 10: ...tM TH8 User Manual Rev A1 1 7MH 018 A11 10 1 4 Block Diagrams 1 4 1 Block Diagram for the tM TH8...
Страница 11: ...tM TH8 User Manual Rev A1 1 7MH 018 A11 11 1 5 Dimensions...
Страница 13: ...tM TH8 User Manual Rev A1 1 7MH 018 A11 13 1 7 Quick Start Please refer to the Quick Start Guide for tM TH8...
Страница 67: ...tM TH8 User Manual Rev A1 1 7MH 018 A11 67 Related Topics Section 1 9 Calibration tM TH8...