background image

 

 

I.A.C.E.R. Srl 

5

 

 

positioned to contact with the skin of the more patient or less next to the zone to be 
treated, to close the circuit.  

The  generator  of  tide  works  within  the  waves  radio  frequency  long  that  oscillate 
between the 0,4 and the 0,5MHZ with a varying power thin to a maximum of 300W 

In  this  way  issue  of  external  energy  is  not  had,  but  there  is  only  a  development  of 
endogenous  or  inside  energy  to  level  of  the  biological  tissue  produced  by  the 
movement of ions and electrolytes. This movement is induced from the strengths of 
attraction and repulsion that produce him among the 2 plates of the condenser. 

Techniques of application  

The  capacitive/resistive  tecartherapy  (which  stands  for  Energy  Transfer  Capacitive 
and  Resistive)  is  a  therapy  that  stimulates  the  body's  natural  repair  processes, 
shortening the time of motor recovery. 

The  diathermy,  through  the  interaction  electromagnetic  energy  /  tissue,  produces  a 
temperature increase that takes place within the tissues in a uniform and controlled 
way.  This  electro-magnetic  interaction  induces  an  ion  flow  with  a  micro-hyperemia 
that  ultimately  promotes  the  release  of  "endogenous  substances"  (mainly  cortisol 
and endorphins) that are used to reduce pain, edema and inflammation. 

I-TECH.AR  stimulates  the  increase  of  blood  flow  in  a  direct  manner,  thanks  to  the 
increase of temperature, and indirectly through the oxygen demand from the treated 
tissues; the increase of blood flow favors the increase of the normal immune system 
activities and stimulates the tissue regeneration.  

I-TECH.AR operates in two modes:  

- in 

capacitive

 mode, it will cause an increase of electrical charge density in proximity 

of the area below the movable electrode and especially at the level of the soft tissues 
such as muscle 

-  in 

resistive

  mode,  the  concentration  of  electrical  charges  and  thus  the  biological 

effect  occurs  in  tissues  with  the  highest  resistance  that  are  between  the  movable 
electrode and the return plate, namely bone, ligaments, etc..  

How it works  

Because the phenomenon of increased charge density occurs, it’s necessary that the 
two capacitor plates are connected to an electric generator potential difference that 
is responsible for supplying charges to the same armor. 

This establishes a real current in accumulation phase goes from the generator to the 
condenser.  As  the  capacitor  builds  up  charge,  the  flow  is  reduced  to  zero  when  the 
capacitor is fully charged. 

After  this  initial  phase  if  the  polarity  of  the  generator  is  reversed,  there  will  be  a 
current in the opposite direction that the load capacitor with opposite polarity to the 
previous.  If  the  generator  reverses  polarity  periodically  there  will  be  a  flow  in  both 
directions, i.e. an alternating current. 

The transfer by 

capacitive contact

 occurs through a capacitive electrode shielded with 

ceramic insulating material (of different sizes depending on the treatment area), that 
mobilize the charges in the subcutaneous tissue.  

The 

resistive  transfer

  occurs  through  an  electrode  resistive  (not  isolated)  that 

mobilizes the charges so that they are concentrated in the areas of greater depth and 
resistivity (bony matrix and deep musculature).  

Capacitive mode acts therefore specifically on the soft tissues (muscles, blood vessels 
and  lymphatics,  adipose  tissue),  the  resistive  mode  acts  instead  on  tissues  with 
greater resistance (bony tissue, cartilage, tendons, serous fascia).  

A  neutral  plate  (return  pole)  is  positioned  in  proximity  of  the  structures  treated  to 
close  the  circuit  applied  according  to  a  geometric  arrangement  that  allows  the 
location  of  the  more  point  resistive  or  capacitive  to  the  affected  area  (e.g.,  the 
popliteal groove in treatment of the knee).  

For a good transfer of the energy to the tissue on the part to be treated a salty gel of 
coupling  is  used  or  however  a  transmission  gel  that  prevents  the  creation  of  an  air 
gap  between  the  electrode  and  the  body  surface  and  face  adhere  better  irradiator 
and epidermis.  

I-TECH.AR  is  a  medical  device  produced  according  to  the  Directive  MED  93/42/CEE 
concerning medical devices.  

I

I

N

N

 

 

G

G

E

E

N

N

E

E

R

R

A

A

L

L

 

 

I.A.C.E.R. srl ha recentemente sviluppato una serie completa di apparati, accessori ed 
attrezzature,  progettati  e  costruiti  secondo  i  più  elevati  standard  qualitativi, 
adottando  tecnologie  d’avanguardia  nel  rispetto  totale  delle  direttive  e  delle  norme 
vigenti. 

Particular attention has been paid to the design, easy operation, function and safety 
of  the  equipment  and  the  final  result  is  this  modern,  compact  unit,  which  offers  an 
extremely logical operative sequence supported by a clearly legible display. 

Содержание I-TECH.AR

Страница 1: ...MNPG112 Rev 2 17 05 2013 USER MANUAL...

Страница 2: ...S UP 7 ACCESSORIES 8 CONNECTIONS 8 DEVICE DESCRIPTION 9 HOW TO USE OF THE DEVICE 11 BEST USE 11 SETTINGS 11 VARIOUS 12 LANGUAGE 13 DEFAULT 13 FREE PROCEDURE 13 LOADING PROGRAMS 14 CREATE PROGRAMS 14 M...

Страница 3: ...ntil the decommissioning of equipment The operating instructions must be available for consultation in the vicinity of the machine and properly stored This manual reflects the state of the art at the...

Страница 4: ...nsider hygienic Italian law T U S 81 2008 on safety in the workplace will not be accepted 3 disassemble accessories and any mechanical supports 4 use original box and packing materials 5 enclose Servi...

Страница 5: ...s aggressive solutions or flammable liquids to clean the external parts of the unit and accessories Using these substances or misusing the accessories will cause the immediate voiding of all warranty...

Страница 6: ...unplug the socket connected to the mains Before every treatment it is recommended to clean with caution all of the accessories and the parts of the equipment that have been to contact with the patien...

Страница 7: ...pplying charges to the same armor This establishes a real current in accumulation phase goes from the generator to the condenser As the capacitor builds up charge the flow is reduced to zero when the...

Страница 8: ...urn plate placed at the popliteal groove Coxarthrosis treatment with resistive electrode positioned on the coxofemoral joint In right or left side decubitus return plate positioned on the inside side...

Страница 9: ...n 1 capacitive probe n 1 kit capacitive electrodes diameters 30mm 50 mm and 70mm n 1 bottle of conductive cream 1000ml n 1 steel plate 1 smart card Check the contents of the package and should any of...

Страница 10: ...for problems or difficulty installation CONNECTIONS The power entry module can be found on the back of the unit and consists of a three pole socket for the cable set an extractible fuse box with two f...

Страница 11: ...I A C E R Srl 9 D DE EV VI IC CE E D DE ES SC CR RI IP PT TI IO ON N Programming panel Color graphic LCD display Rear panel Connector for probes Encoder control knob Connector for plate...

Страница 12: ...ive and resistive probes Capacitive electrodes Resistive electrodes Ground connection Integrated power module Ventola di aerazione Connector for STEEL PLATE Connectors for PROBES Audio connector not u...

Страница 13: ...alised application BEST USE After having installed and correctly positioned the machine as per the instructions described in the previous sections and connecting the applicator correctly plug the mach...

Страница 14: ...ormatted You can also use the FORMAT SMART CARD function to delete it completely If you do this it can be used on a different unit Select one of the memory supports to format turning to the right the...

Страница 15: ...o the handle according to its diameter Per modificare tali parametri procedere come segue 1 turn the encoder knob to select the parameter 2 press the knob to highlight the parameter 3 rotate the knob...

Страница 16: ...TART button Before initiation of therapy however you can modify any parameter as discussed in section DEFAULT SETTINGS and FREE PROCEDURE but the program can neither be stored nor renamed C CR RE EA A...

Страница 17: ...denatured alcohol Store with care the probes at the conclusion of every treatment Contact authorised dealers for information regarding original spare parts or components Do not spray or pour liquid on...

Страница 18: ...ocopiers fax machines etc or any electric or electronic equipment used in these environments as long as said equipment complies with the ELECTROMAGNETIC COMPATIBILITY directive In any case in order to...

Страница 19: ...3 15 A T 5 x 20 mm 115 Vac 6 3 A T 5 x 20 mm Peak power 200 Wpk max Peak voltage Capacitive use 300 V max Resistive use 105 V max Probes supplied Resistive probe Capacitive probe Emission Frequency of...

Страница 20: ...oduct shall not be disposed of as domestic waste The user must dispose of scrap equipment by taking it to a recognised centre of recycling of electrical and electronic equipment Appendix B LABELS Symb...

Страница 21: ...min res 15 Cervicalgy x Medium livel 50 70 10 15 Cervical brachialgia x Medium livel 70 10 15 Artropatie da malattie autoimmuni x Medium livel I 12 20 Epicondylitis x Medium livel 50 12 15 Gonarthros...

Страница 22: ...ectrostatic discharge ESD IEC 61000 4 2 6kV contact 6kV contact Floors should be wood concrete or ceramic tile If floors are covered with synthetic material the relative humidity should be at least 30...

Страница 23: ...th accuracy To assess the electromagnetic environment due to fixed RF transmitters an electromagnetic site survey should be considered If the measured field strength in the location in which the ME EQ...

Страница 24: ...nd warehouse 30030 MARTELLAGO VE Via S Pertini 24 A Tel 0039 041 5401356 Fax 0039 041 5402684 Cod Fisc P IVA IT 00185480274 R E A VE N 120250 M VE001767 e mail iacer iacer it Internet www itechmedical...

Отзывы: