I.A.C.E.R. Srl 

14

 

 

If you press the STOP button the emission finally ends and you return to the screen in 
fig.8. 

Pressing the EXIT button returns you to the screen in fig.2. 

 

 

Fig.9 

Fig.10 

L

L

O

O

A

A

D

D

I

I

N

N

G

G

 

 

P

P

R

R

O

O

G

G

R

R

A

A

M

M

S

S

 

 

Pressing  the  LOAD  PROGRAM  button  on  the  screen  (see  fig.2),  appears  the  list  of 
therapeutic protocols stored in the main memory (as shown by the frame around the 
STANDARD PROGRAMS button which is selected by default). These programs cannot 
be  deleted  but  can  be  overwritten  by  changing  the  parameters  of  interest  without 
saving. 

Instead  pressing  the  USER  PROGRAMS  button,

 

appear  on  the  screen  the  numbered 

sections  (with  default  parameters)  that  will  contain  the  programs  created  with  the 
CREATE PROGRAMS function, and you can load programs stored in the user memory. 

Finally  pressing  the  SMART-CARD  PROGRAMS  taste,  appear  on  the  screen  the 
numbered sections (with default parameters) that will contain the programs created 
with the CREATE PROGRAMS function, and you can load customized programs, stored 
in the smart-card.  

NOTE: 

If  you  save  a  program  on  Smart-Card  but  the  Smart-Card  is  not  inserted,  the 

operator is alerted by an alarm message to the not inclusion of the Smart Card into 
place  and  then  the  inability  to  be  able  to  continue  in  the  operation  of  storage 
programs.  

The stored programmes reflect the fruit of many years experience supporting expert 
professional operators. Appendix 

C

 shows a list of the programmes available . 

Pressing the EXIT button (whatever the memory selected) returns to the screen of the 
fig.2.  

To start the desired treatment, turn the encoder knob to reach the desired protocol, 
then press it to confirm the selection.  

Once  the  display  shows  the  selected  program  screen,  you  can  go  directly  to  its 
execution by simply pressing the START button.  

Before initiation of therapy, however, you can modify any parameter, as discussed in 
section  DEFAULT  SETTINGS  and  FREE  PROCEDURE,  but  the  program  can  neither  be 
stored nor renamed. 

C

C

R

R

E

E

A

A

T

T

E

E

 

 

P

P

R

R

O

O

G

G

R

R

A

A

M

M

S

S

 

 

 

 

This function allows to store “customized” therapeutic programs in the Smart-card  or 
in the user memory, which are the only memory available to save the new programs. 

Pressing the CREATE PROGRAMS taste on the screen (see fig.2) to create a program; 
appears the screen of the fig.11.  

Pressing  the  encoder  knob, you  can  start  by  default  the  creation  of  the  program  on 
the user memory (as shown by the frame around the USER PROGRAM button).  

Instead,  press  the  SMART-CARD  PROGRAMS  button  to  create  a  program  on  the 
smart-card.   

NOTE: 

If  you  save  a  program  on  Smart-Card  but  the  Smart-Card  is  not  inserted,  the 

operator is alerted by an alarm message (see dfig.12) to the not inclusion of the Smart 
Card  into  place  and  then  the  inability  to  be  able  to  continue  in  the  operation  of 
storage programs. 

Once  you  select  the  support  of  memory  in  which  to  save  the  program,  press  the 
encoder knob to confirm the selection. It appears the screen of the figure 13. 

At  this  screen,  to  assign  a  name  to  the  program  press  the  encoder  knob:  appears  a 
cursor  under  the  first  character  (see  Figure  14),  indicating  the  possibility  to  switch 
between the characters that you want to change by turning the knob. Then press the 
encoder knob at the character to confirm your choice. 

Now the selected character is surrounded by two sliders (fig. 15), which means that 
the  character  is  changed.  Rotating  the  encoder  to  choose  a  new  character  to  enter 
and  press  the  encoder  knob  to  confirm  the  selection.  This  exits  from  the  routine  to 
change the selected character.  

Stop

Start

Exit

Potenza        0%

Tempo: 05:00

Resistivo

Tempo     15min 

Manipolo  30min 

DEFAULT

Tempo     15min 

Modalità:sequenza 

Capacitivo

Manipolo  30min 

PAUSE

Time     15min 

Modality:sequence 

Time     15min 

Probe    30min 

Probe    30min 

DEFAULT

Capacitive

Resistive

Содержание I-TECH.AR

Страница 1: ...MNPG112 Rev 2 17 05 2013 USER MANUAL...

Страница 2: ...S UP 7 ACCESSORIES 8 CONNECTIONS 8 DEVICE DESCRIPTION 9 HOW TO USE OF THE DEVICE 11 BEST USE 11 SETTINGS 11 VARIOUS 12 LANGUAGE 13 DEFAULT 13 FREE PROCEDURE 13 LOADING PROGRAMS 14 CREATE PROGRAMS 14 M...

Страница 3: ...ntil the decommissioning of equipment The operating instructions must be available for consultation in the vicinity of the machine and properly stored This manual reflects the state of the art at the...

Страница 4: ...nsider hygienic Italian law T U S 81 2008 on safety in the workplace will not be accepted 3 disassemble accessories and any mechanical supports 4 use original box and packing materials 5 enclose Servi...

Страница 5: ...s aggressive solutions or flammable liquids to clean the external parts of the unit and accessories Using these substances or misusing the accessories will cause the immediate voiding of all warranty...

Страница 6: ...unplug the socket connected to the mains Before every treatment it is recommended to clean with caution all of the accessories and the parts of the equipment that have been to contact with the patien...

Страница 7: ...pplying charges to the same armor This establishes a real current in accumulation phase goes from the generator to the condenser As the capacitor builds up charge the flow is reduced to zero when the...

Страница 8: ...urn plate placed at the popliteal groove Coxarthrosis treatment with resistive electrode positioned on the coxofemoral joint In right or left side decubitus return plate positioned on the inside side...

Страница 9: ...n 1 capacitive probe n 1 kit capacitive electrodes diameters 30mm 50 mm and 70mm n 1 bottle of conductive cream 1000ml n 1 steel plate 1 smart card Check the contents of the package and should any of...

Страница 10: ...for problems or difficulty installation CONNECTIONS The power entry module can be found on the back of the unit and consists of a three pole socket for the cable set an extractible fuse box with two f...

Страница 11: ...I A C E R Srl 9 D DE EV VI IC CE E D DE ES SC CR RI IP PT TI IO ON N Programming panel Color graphic LCD display Rear panel Connector for probes Encoder control knob Connector for plate...

Страница 12: ...ive and resistive probes Capacitive electrodes Resistive electrodes Ground connection Integrated power module Ventola di aerazione Connector for STEEL PLATE Connectors for PROBES Audio connector not u...

Страница 13: ...alised application BEST USE After having installed and correctly positioned the machine as per the instructions described in the previous sections and connecting the applicator correctly plug the mach...

Страница 14: ...ormatted You can also use the FORMAT SMART CARD function to delete it completely If you do this it can be used on a different unit Select one of the memory supports to format turning to the right the...

Страница 15: ...o the handle according to its diameter Per modificare tali parametri procedere come segue 1 turn the encoder knob to select the parameter 2 press the knob to highlight the parameter 3 rotate the knob...

Страница 16: ...TART button Before initiation of therapy however you can modify any parameter as discussed in section DEFAULT SETTINGS and FREE PROCEDURE but the program can neither be stored nor renamed C CR RE EA A...

Страница 17: ...denatured alcohol Store with care the probes at the conclusion of every treatment Contact authorised dealers for information regarding original spare parts or components Do not spray or pour liquid on...

Страница 18: ...ocopiers fax machines etc or any electric or electronic equipment used in these environments as long as said equipment complies with the ELECTROMAGNETIC COMPATIBILITY directive In any case in order to...

Страница 19: ...3 15 A T 5 x 20 mm 115 Vac 6 3 A T 5 x 20 mm Peak power 200 Wpk max Peak voltage Capacitive use 300 V max Resistive use 105 V max Probes supplied Resistive probe Capacitive probe Emission Frequency of...

Страница 20: ...oduct shall not be disposed of as domestic waste The user must dispose of scrap equipment by taking it to a recognised centre of recycling of electrical and electronic equipment Appendix B LABELS Symb...

Страница 21: ...min res 15 Cervicalgy x Medium livel 50 70 10 15 Cervical brachialgia x Medium livel 70 10 15 Artropatie da malattie autoimmuni x Medium livel I 12 20 Epicondylitis x Medium livel 50 12 15 Gonarthros...

Страница 22: ...ectrostatic discharge ESD IEC 61000 4 2 6kV contact 6kV contact Floors should be wood concrete or ceramic tile If floors are covered with synthetic material the relative humidity should be at least 30...

Страница 23: ...th accuracy To assess the electromagnetic environment due to fixed RF transmitters an electromagnetic site survey should be considered If the measured field strength in the location in which the ME EQ...

Страница 24: ...nd warehouse 30030 MARTELLAGO VE Via S Pertini 24 A Tel 0039 041 5401356 Fax 0039 041 5402684 Cod Fisc P IVA IT 00185480274 R E A VE N 120250 M VE001767 e mail iacer iacer it Internet www itechmedical...

Отзывы: