HOCT-1/1F
6.5.
Anterior segment image operation (optional)
6.5.1.
Preparation for anterior segment operation
1.
Check the lens surface of anterior segment adapter is clean.
2.
Thread anterior segment adapter (A) to objective lens holder (B), and check there is no tilting or misalignment
of adopter. Don’t lock it too hard.
3.
Check anterior headrest rubber is clean.
4.
Hang anterior headrest rubber to headrest of instrument (C) first, then press anterior headrest rubber to be
fixed firmly.
5.
Check the mounted anterior headrest rubber does not incline.
6.5.2.
Capturing anterior segment
1.
The first procedure is same as ‘6.4 General Operation: procedure 2 ~ 7’.
•
On the anterior mode, function of auto shooting and auto tracking is not available.
2.
Move the body to align patient’s eye. Move the body slowly while watching patient’s eye and body, because
working distance is just 15mm that the front lens is very close to patient's eye.
Содержание HOCT-1
Страница 1: ...OPTICAL COHERENCE TOMOGRAPHY HOCT 1 1F USER MANUAL...
Страница 77: ...HOCT 1 1F 3 Select the TRANSFER icon to send the report to the DICOM server...
Страница 102: ...102 1 Selecting PRINT icon shows printer option window 2 Select the Save icon the Select Storage Location window appears...
Страница 109: ...HOCT 1 1F 11 Selecting FULL Screen icon shows the current Bscan image in full screen...
Страница 118: ...118 9 Selecting FULL Screen icon shows the current Bscan image in full screen...
Страница 131: ...HOCT 1 1F Vessel Faz...
Страница 149: ...HOCT 1 1F 8 3 Drawings of System...