Scan along 12 lines for cornea.
<Scan Options>
•
Range: Choose scan range from 6mm, 4.5mm, 9mm, or 12mm.
•
AScan Points: Choose the number of A-scans that constitute one B-scan. Possible options are 1024,
512, or 256.
•
Bscan Lines: Choose the number of B-scans that constitute one measurement. Possible options
depend on the number of A-scans set.
•
Overlap: Choose the number of B-scans to be averaged to produce one B-scan image. Possible
options are enumerated.
•
Direction: Set a direction which is horizontal or vertical.
•
Space: Set an interval between neighboring scan lines.
•
Sensitivity: Choose scan quality from Normal, Fine, or Ultra fine. The latter produce better image but
take longer.
-
Scan Overlap: Increase image resolution by overlapping multiple Bscan images.
Bscan images with different number of overlapping images.
< Overlap : 1 >
< Overlap : 15 >
< Overlap : 30 >
< Overlap : 45 >
-
Scan Range: Select the size of area to be measured.
Bscan images of each Scan Ranges
Macular Line Analysis Screen
Macular 3D Analysis Screen
Содержание HOCT-1
Страница 1: ...OPTICAL COHERENCE TOMOGRAPHY HOCT 1 1F USER MANUAL...
Страница 77: ...HOCT 1 1F 3 Select the TRANSFER icon to send the report to the DICOM server...
Страница 102: ...102 1 Selecting PRINT icon shows printer option window 2 Select the Save icon the Select Storage Location window appears...
Страница 109: ...HOCT 1 1F 11 Selecting FULL Screen icon shows the current Bscan image in full screen...
Страница 118: ...118 9 Selecting FULL Screen icon shows the current Bscan image in full screen...
Страница 131: ...HOCT 1 1F Vessel Faz...
Страница 149: ...HOCT 1 1F 8 3 Drawings of System...