HOCT-1/1F
A.
Try moving internal fixation target position by pressing fixation icon (
) and changing
position of green cross (
) if needed.
When green cross position changes, the position of internal fixation target is also changed.
B.
Try to change scan position by dragging scan range while scan range icon turned on. If reset
scan position icon (
) is pressed, scan position moves to the default center position.
11.
Capture image and check image quality (in Fundus mode).
(1)
Change focus position by pressing optimize icon on the screen or optimize button on the body (A).
(2)
Press the button on joystick to capture image.
If auto shooting is on, ‘11-2. Optimize’ and ‘11.3 Capture’ is accomplished automatically.
(3)
Check the result and decide to store or discard and retry.
①
If the image is satisfactory, press OK icon (
) to save image.
②
If the image is not satisfactory, press retry icon (
) and retry image capturing.
A.
If fundus image result too bright or too dark because of lighting, regulate the flash intensity using
flash icon (
) in observation mode.
B.
If fundus image is too dark because of small pupil size of patient, try small pupil mode by using
small pupil icon (
) in observation mode.
Содержание HOCT-1
Страница 1: ...OPTICAL COHERENCE TOMOGRAPHY HOCT 1 1F USER MANUAL...
Страница 77: ...HOCT 1 1F 3 Select the TRANSFER icon to send the report to the DICOM server...
Страница 102: ...102 1 Selecting PRINT icon shows printer option window 2 Select the Save icon the Select Storage Location window appears...
Страница 109: ...HOCT 1 1F 11 Selecting FULL Screen icon shows the current Bscan image in full screen...
Страница 118: ...118 9 Selecting FULL Screen icon shows the current Bscan image in full screen...
Страница 131: ...HOCT 1 1F Vessel Faz...
Страница 149: ...HOCT 1 1F 8 3 Drawings of System...