148
8.2.
Specifications
OCT (HOCT-1)
Principle
Spectral domain OCT, Fundus digital photography
Light source
840 nm
Scan speed
Max. 68,000 A-scan/sec.
Resolution in tissue
20 um(Lateral), 6~7 um(z-axis) at index 1.36
Scan Range
X: 6 ~ 12 mm, Y: 6 ~ 9 mm, Z: 2.34 mm
Display resolution
X: 5.85 um, Y: 23.40 um, Z: 3.05 um
Minimum pupil diameter
2.5 mm
Scan patterns
Macular: Macular Line, Macular Cross, Macular Radial,
Macular 3D, Macular Raster
Disk: Disc Circle, Disc Radial, Disc 3D, Disc Raster
Optical power at cornea
≤ 650 uW
Acquisition time of 3D image
1.4 sec. (Normal mode, A512 x B96)
Depth Accuracy
(measuring 1mm glass)
±3%
Angiography Range
X:3 ~ 12mm, Y:3~9mm (Optional)
Angiography Analysis
FAZ, Vessel density Tools (Optional)
Fundus Camera (HOCT-1F Only)
Type
Non-mydriatic fundus camera
Resolution
60 line pair/mm (center), 40 line pair/mm (middle), 25 line
pair/mm (periphery)
Angle of view
45˚
Camera
Built-in 20Mega pixel, Color
Minimum pupil diameter
4.0 mm (Normal mode), 3.3 mm (Small pupil mode)
Flash light
White light, 10 levels
Pixel pitch at fundus
3.69 um
Common specification
Working distance
33 mm
Display
12.1 inch, 1280x800 pixel, Touch panel color LCD
Dioptric compensation for
patient’s eye
-
33D ~ +33D total
-
13D ~ +13D with no compensation lens
+7D ~ +33D with plus compensation lens
-
33D ~ -7D with minus compensation lens
Internal fixation target
LCD (internal), White LED (external)
Fundus illumination light
760 nm
Horizontal movement
70 mm (back and forth), 100 mm (left and right)
Vertical movement
30 mm
Chinrest movement
62 mm (up and down), motorized
Auto tracking
30mm (up and down), 10 mm (right and left), 10mm (back
and forth)
Power supply
AC 100 - 240 V, 50/60 Hz, 1.6 - 0.7 A
PC
Built in computer
LCD Tilting Angle
70˚
External port
2 USB, 1 DP, 1 RGB, 2 LAN
Dimensions
330(W) x 542(D) x 521(H) mm
Mass
30 kg
Anterior segment adapter (optional)
Working distance
15 mm
Scan range
6 ~ 9 mm (width), 2.3 mm (depth)
Scan pattern
ACA line, Anterior Radial
Mass
30 kg
Wide Anterior segment adapter (optional)
Working distance
15 mm
Scan range
16 mm (width), 2.3 mm (depth)
Scan pattern
ACA line, Anterior Radial
Содержание HOCT-1
Страница 1: ...OPTICAL COHERENCE TOMOGRAPHY HOCT 1 1F USER MANUAL...
Страница 77: ...HOCT 1 1F 3 Select the TRANSFER icon to send the report to the DICOM server...
Страница 102: ...102 1 Selecting PRINT icon shows printer option window 2 Select the Save icon the Select Storage Location window appears...
Страница 109: ...HOCT 1 1F 11 Selecting FULL Screen icon shows the current Bscan image in full screen...
Страница 118: ...118 9 Selecting FULL Screen icon shows the current Bscan image in full screen...
Страница 131: ...HOCT 1 1F Vessel Faz...
Страница 149: ...HOCT 1 1F 8 3 Drawings of System...