Fetal ECG input
Direct FECG protection category
BF
Scalp: minimum signal threshold
30
µ
V
Patient leakage current
100
µ
A max (240V)
Sampling rate
± 1 ms
Heart rate
Calculated to ± 0.25 bpm
Maternal ECG input
Direct FECG protection category
BF
Scalp: minimum signal threshold
30
µ
V
Patient leakage current
100
µ
A max (240V)
Sampling rate
± 1 ms
Heart rate
Calculated to ± 0.25 bpm
Contractions (internal IUP)
Protection category
BF
Input connections
Fully isolated
Pressure range (IUP)
0-100 mmHg & 0-15.0 kPa
Breakdown voltage
2.5kV rms to ground.
Transducer sensitivity
5 mV/V/mmHg
IUP Accuracy
± 5% (nominal transducer)