Sonicaid Team Operator’s Manual
22
1.8
Team printer to Team base unit assembly
The Team base unit is supplied already assembled to the Team printer.
To disassemble
1 Press the release button beneath the left edge of the printer platen.
2 Lift the platen to the left and off the top of the printer.
3 Remove the paper pack.
4 Using a screwdriver, release the centre fixing screw (approximately 4 turns).
5 Remove the printer unit from the main unit.
6 Re-fit the paper pack and platen.
To reassemble
1 Remove the centre blanking-plug (if fitted) from the Team base unit top.
2 Position the printer unit on top of the base unit. The feet of the printer unit will
locate in depressions on the base unit top.
3 Remove the platen and lift the paper pack for access to the screw-head beneath.
4 Using a screwdriver, push down and secure the screw with approximately 4 turns.
5 Re-fit the paper pack and platen.
1 Printer platen shown removed
2 Centre fixing screw
3 Release button
Содержание SonicaidTeam
Страница 14: ...Sonicaid Team Operator s Manual 14 Addresses...
Страница 68: ...Sonicaid Team Operator s Manual 68 Example printout of TeamCare trace and analysis report...
Страница 69: ...Sonicaid Team Operator s Manual 69 Example printout of TeamCare twins analysis results...