Sonicaid Team Operator’s Manual
54
4.4
Alarms
Alarm setup menu
1 Press [MENU] once.
2 Select [ALARM].
< < < S I G N AL L O S S
< < < L O W F H R
F E T A L M O V E M E N T > > >
<<<H IG H F H R
E X IT >>>
Signal loss alarm
You can set a signal loss threshold above which an alarm will occur. This threshold is
a percentage of the last 5-minute period. With a setting of 20%, for example, one-
minute of signal loss in 5 minutes will trigger the alarm. This alarm is only active
when the printer or store is active.
To set a signal loss alarm:
1 From the Alarm setup menu press [SIGNAL LOSS].
2 Select [ALARM ON SILENT] or [ALARM ON AUDIBLE].
If you set a silent alarm, and the alarm is triggered, a notification appears on the
display message bar, but the Team does not beep.
3 Enter the % signal loss required.
4 Team asks IS THIS CORRECT? If it is correct, press [ACCEPT]. If it is not correct, press
[RE-ENTER].
To turn the signal loss alarm off:
1 From the Alarm setup menu press [SIGNAL LOSS].
2 Select [ALARM OFF].
Low and high FHR alarms
You can set fetal heart rate thresholds such that an alarm occurs if the signal remains
above or below the threshold for a specified time (known as the ‘delay time’). These
alarms are only active when the printer or store is active.
Содержание SonicaidTeam
Страница 14: ...Sonicaid Team Operator s Manual 14 Addresses...
Страница 68: ...Sonicaid Team Operator s Manual 68 Example printout of TeamCare trace and analysis report...
Страница 69: ...Sonicaid Team Operator s Manual 69 Example printout of TeamCare twins analysis results...
Страница 79: ...Sonicaid Team Operator s Manual 79 Printout showing Team Trend analysis results...