Sonicaid Team Operator’s Manual
30
Using the printer
1 To switch the printer on, press the button on the printer front panel.
2 To fast forward the paper, press and hold down the printer button.
3 To stop the printer, press the printer button again.
Using the second ultrasound transducer for twins
1 Connect the second ultrasound transducer (blue) to the Team. The display
switches to twin heart rate display.
2 Place both ultrasound transducers on the patient’s abdomen in the optimum
position. Use the blue ultrasound transducer to monitor the first, presenting twin.
3 Make sure each fetal heart rate is from a separate fetus. See Section 3.4. If in doubt,
ask for assistance. Secure the ultrasound transducers with belts and buckles.
4 To select Audio, press the bottom left button on the keypad. The active audio
channel is highlighted on the display.
Monitoring fetal ECG
Using a Sonicaid scalp electrode:
1 Put electrode gel on the base of the leg plate, then strap the leg plate to the
patient’s thigh. Secure with the belt.
2 Connect the FECG lead to the Team.
3 Once the membranes are ruptured, attach the electrode to the fetus as described
in the electrode instructions.
4 Connect the electrode leads to the leg plate. Make sure a good signal is
maintained.
5 Wait for the signal to stabilize and a clear fetal heart rate to be displayed on the
Team base unit display. Then adjust the volume control.
Using a Safelinc electrode:
1 Attach the FECG lead to the mother’s leg.
2 Once the membranes are ruptured, attach the FECG electrode to the fetal pre-
senting part.
3 Connext the FECG electrode to the FECG lead.
4 Wait for the signal to stabilize and a clear fetal heart rate to be displayed on the
Team base unit display. Then adjust the volume control.
See also Section 3.3.
Содержание SonicaidTeam
Страница 14: ...Sonicaid Team Operator s Manual 14 Addresses...
Страница 68: ...Sonicaid Team Operator s Manual 68 Example printout of TeamCare trace and analysis report...
Страница 69: ...Sonicaid Team Operator s Manual 69 Example printout of TeamCare twins analysis results...
Страница 79: ...Sonicaid Team Operator s Manual 79 Printout showing Team Trend analysis results...