Sonicaid Team Operator’s Manual
80
7.5
Viewing trend data
To see a trend of the analysis results for up to the last four hours, press [TREND] in
the results display screen.
The Trend display screen returns to the fetal heart rate display after two minutes, or
when you press [EXIT].
T I M E
2 1 : 0 6
2 2 : 0 6
2 3 : 0 6
0 0 : 0 6
6 0 M I N B A S E L I N E
1 3 9
1 4 1
1 4 3
1 3 8
6 0 M I N S T V
7 . 8
8 . 6
8 . 1
8 . 0
6 0 M I N D E C E L
2 8 0
2 9 0
3 1 0
3 5 0
C O N F I D EN C E
M
H
H
H
U L T - Y
EX I T >>>
For twins, the results shown are for the fetus on the currently selected audio channel
(ULT-Y or ULT-B/FECG). If no audio channel is selected, the default is ULT-Y.
7.6
Analysis parameters and calculations
The baseline
The analysis fits a baseline to the fetal heart rate trace through data points. Data
points are pulse intervals averaged over 1/16 minute. The baseline is fitted in such a
way as to follow slow, but not rapid, changes in the fetal heart rate.
The baseline is re-fitted at every analysis, as further information becomes available.
Confidence indicator
The confidence measure for baseline fit is based on points for possible errors:
Points
Confidence
0 or 1
High
2 or 3
Medium
4 +
Low
Note
: the confidence indicator defaults to Low for the first analysis at 15 minutes.
Содержание SonicaidTeam
Страница 14: ...Sonicaid Team Operator s Manual 14 Addresses...
Страница 68: ...Sonicaid Team Operator s Manual 68 Example printout of TeamCare trace and analysis report...
Страница 69: ...Sonicaid Team Operator s Manual 69 Example printout of TeamCare twins analysis results...
Страница 79: ...Sonicaid Team Operator s Manual 79 Printout showing Team Trend analysis results...