
Sonicaid Team Operator’s Manual
106
Appendix 2: Transducer Problems
The following tests will show whether there is a problem with an Ultrasound
transducer. If there is a problem, contact the Service Department of Sonicaid
Products, or their appointed service agent.
Preliminary
1 Connect the ultrasound transducer to the Team.
2 Turn on the Team.
3 Select the required audio channel.
4 Adjust the volume to the required level.
System test
1 Hold the ultrasound transducer in one hand, with the transducer face against the
palm.
2 Stroke the back of the hand repeatedly with one finger. See diagrams below.
If necessary, use water or gel to obtain good contact between the palm and
transducer.
3 Check that the audio output, pulse lamp, heart rate display and printer trace on
the Team are synchronised with the finger movement.
Содержание SonicaidTeam
Страница 14: ...Sonicaid Team Operator s Manual 14 Addresses...
Страница 68: ...Sonicaid Team Operator s Manual 68 Example printout of TeamCare trace and analysis report...
Страница 69: ...Sonicaid Team Operator s Manual 69 Example printout of TeamCare twins analysis results...
Страница 79: ...Sonicaid Team Operator s Manual 79 Printout showing Team Trend analysis results...