Sonicaid Team Operator’s Manual
64
Actogram
Event marks recorded by the Actogram facility are not used by the Team Care
analysis.
Alarms
During analysis the Signal Loss alarm is fixed at 30%. In addition, there is a fixed
Toco alarm that alerts the user to a constant Toco value for 10 minutes. Once this
alarm has been acknowledged, it will not re-alarm during the same analysis.
Fetal ECG mode
Since the analysis is not valid during labour, it does not run on the secondary channel
if this is using a fetal ECG scalp electrode.
6.5
Using the analysis
Starting the analysis
1 Set up Team as you would to record a normal trace.
2 Press [MENU] once. Check that the display says ANALYSIS IS ON. If it says
ANALYSIS IS OFF, then select [TURN ANALYSIS ON].
3 Press the printer button once to start the printer.
4 Team beeps, and asks you to enter the gestational age of the fetus.
5 Enter the gestational age.
6 The Team stops beeping, and the printer starts.
Stopping the analysis
1 Press the printer button once to stop the printer.
2 The trace fast forwards, and the analysis results are printed.
If you stop the printer before 10 minutes, Team says CANNOT ANALYSE. LESS THAN
10 MINUTES DATA, and asks CONTINUE PRINT & ANALYSIS?
Select either [HALT PRINTER] or [CONTINUE PRINT].
Содержание SonicaidTeam
Страница 14: ...Sonicaid Team Operator s Manual 14 Addresses...
Страница 68: ...Sonicaid Team Operator s Manual 68 Example printout of TeamCare trace and analysis report...
Страница 69: ...Sonicaid Team Operator s Manual 69 Example printout of TeamCare twins analysis results...
Страница 79: ...Sonicaid Team Operator s Manual 79 Printout showing Team Trend analysis results...