26
*3-wire: Assumes matched leads with a total resistance not exceeding 100Ω.
Resolution (0~1000)Ω: 0.01Ω;
(1.0~3.2)kΩ: 0.1Ω.
Source
Range
Excitation Current
from Measurement
Device
Accuracy
(15~400)Ω
(0.15~0.5)mA
±0.15Ω
(15~400)Ω
(0.5~3)mA
±0.1Ω
(0.4~1.5)kΩ
(0.05~0.8)mA
±0.5Ω
(1.5~3.2)kΩ
(0.05~0.4)mA
±1.0Ω
Clear of resistance before sourcing according to ‘8.4 Clear of Resistance and RTDs’.
Frequency
Range
Accuracy
Measurement
(1~1100)Hz
±0.05%
(1.0~10.0)kHz
±0.05%
Sensitivity: Not less than 1V (peak- peak)
Wave form: Square wave
Source
(1~1100)Hz
±0.05%
(1.0~5.0)kHz
±0.05%
Wave form: (0~8)V (peak-peak)
Load drive capability: 3mA
DC Current
Range
Accuracy
Measurement
(0~24)mADC
±0.02%
Source
(0~24)mADC
±0.02%
Thermocouple
(TC)
Type
Range
Measure and Source Accuracy
(When Cold Junction temperature is 0
℃
)
J
(-200~0)°C
±0.8°C
(0~1200)°C
±0.5°C
K
(-200~0)°C
±1.0°C
(0~1370)°C
±0.6°C
T
(-200~0)°C
±1.0°C
(0~400)°C
±0.6°C
E
(-100~0)°C
±0.7°C
(0~950)°C
±0.5°C
R
(-20~0)°C
±2.3°C
(0~500)°C
±1.6°C
(500~1750)°C
±1.2°C
S
(-20~0)°C
±2.3°C
(0~500)°C
±1.6°C
Содержание EDEKA TC-200