22
jacks and press right
.
Clear of Measuring mV DC in lower display:
Press
until the upper display shows “mV”. Short the test leads with
Source/Measure jacks and press right
.
8.2 Clear of Sourcing Voltage
To clear offset of sourcing V DC, proceed as follows:
(1)
Clear of V DC in upper display according to ‘10.1 Clear of Measuring Voltage’.
(2)
Press
to select Source mode.
(3)
Press
until the lower display shows “V”.
(4)
Connect the EDEKA TC-200 Multi-function process calibrator’s voltage output to its
voltage input according to the flashing jacks as Figure3- 1 shows.
(5)
Press right
after the value is stable.
To clear offset of sourcing mV DC, proceed as follows:
(1)
Clear of mV DC in upper display according to ‘10.1 Clear of Measuring Voltage’.
(2)
Press
to select Source mode.
(3)
Press
until the lower display shows “mV”.
(4)
Connect the EDEKA TC-200 Multi-function process calibrator’s voltage output to its
voltage input according to the flashing jacks as Figure3- 1 shows.
(5)
Press right
after the value is stable.
8.3 Clear of Measuring with TC
To clear offset of measuring temperature with TC in the upper display, proceed as follows:
(1)
Press left
to select the decided TC.
(2)
Short the test leads with Measure jacks.
(3)
Press left
after the value is stable.
To clear offset of measuring temperature with TC in the lower display, proceed as follows: