4.3 Sourcing Resistance
To select a resistance sourcing function, proceed as follows:
(1)
Press
until the lower display shows “SOURCE”.
(2)
Press
to select resistance until the display shows “Ω”.
(3)
Connect the test leads as shown in Figure6- 3.
(4)
Enter the desired output value by pressing numeric keypads and press
.
CAUTION: Clear of resistance before sourcing according to ‘8.4 Clear of Resistance and RTD’.
4.4 Sourcing Frequency
The function of Sourcing Frequency is only applicable to EDEKA TC-200 Multi-function
process calibrator. To select a frequency sourcing function, proceed as follows:
(1)
Press
until the lower display shows “SOURCE”.
(2)
Press
to select frequency until the display shows “Hz”.
(3)
Connect the test leads as shown in Figure6- 4.
(4)
Enter the desired output value by pressing numeric keypads and press
.
Amplitude of the output frequency is displayed at the bottom right corner. You can press
to adjust the amplitude. The step of amplitude is 0.5V, and the range is (0-8)V.
Figure6- 3 Resistance Sourcing Connections