3.9 Measuring Temperature with RTDs
The EDEKA TC-200 Multi-function process calibrator accepts RTD types shown in
Table5- 2. RTDs are characterized by their resistance at 0°C, which is called the “ice
point” or R
0
. The most common R
0
is 100Ω. The EDEKA TC-200 Multi-function process
calibrator accepts RTD measurement input in 2-, 3-, or 4-wire connections, with the
3-wire connection the most common. Change the unit of temperature according to ‘8
Switching Temperature Unit’.
To measure temperature using an RTD input, proceed as follows:
(1)
Press
until the lower display shows “MEASURE”.
(2)
Press
to select the desired RTD type.
(3)
Press
to select a 2-, 3-, or 4-wire mode. Connect the test leads as shown in
Figure5- 9.
CAUTION: Clear of RTD before measuring, according to ‘8.4 Clear of Resistance and RTD’.
Table5- 1 Thermocouple Types Accepted
Type
Specified Range
J
(-200~1200)°C
K
(-200~1370)°C
T
(-200~400)°C
E
(-100~950)°C
R
(-20~1750)°C
S
(-20~1750)°C
B
(600~1800)°C
N
(-200~1300)°C
Table5- 2 RTD Types Accepted
Type
Ice Point
(
R
0
)
α
Range
Cu50
50Ω
0.00428Ω/°C
(-50~150)°C
Pt100(385)
100Ω
0.00385Ω/°C
(-200~800)°C
Pt100(3916)
100Ω
0.003916Ω/°C
(-200~510)°C
Pt200(385)
200Ω
0.00385Ω/°C
(-200~630)°C
Pt500(385)
500Ω
0.00385Ω/°C
(-200~630)°C
Pt1000(385)
1000Ω
0.00385Ω/°C
(-200~630)°C