7
You should see the measurement readings in the upper display.
3.3 Measure Current With Loop Power
(1)
Press
until the upper display shows “MEASURE LOOP”.
(2)
Connect the test leads to the transmitter current loop jacks as shown in Figure5- 3.
(3)
The EDEKA TC-200 Multi-function process calibrator supplies an internal 24V loop
power.
You should see the measurement readings in the upper display.
CAUTION: Never short two test leads in “MEASURE LOOP” mode.
Figure5- 1 Measuring Voltage (Upper Display)
Figure5- 2 Measuring Current (Upper Display)
Содержание EDEKA TC-200
Страница 1: ...EDEKA TC 200 Temperature Calibrator Huazheng Electric Manufacturing Baoding Co Ltd...
Страница 2: ......
Страница 6: ......