16
Figure6- 5 Connections for Simulating a Thermocouple
Figure6- 6 Connections for Simulating an RTD
4.7 Ramping and Stepping the Output
Two ways are available for adjusting the value of source function.
Stepping the output manually with
or numeric keypads;
Auto ramping the output.
Auto ramping can be applied in all sourcing functions, which allow you to source a varying
output from the EDEKA TC-200 Multi-function process calibrator continuously.
When you press
, the EDEKA TC-200 Multi-function process calibrator produces a
continuously repeating 0% ~ 100% ~ 0% ramp in your choice of three ramp waveforms:
0%~100%~0% 40 seconds smooth ramp;
0%~100%~0% 10 seconds smooth ramp;
0%
-
25%
-
50%
-
75%
-
100%
-
0%, stair-step ramp in 25% steps, pausing 25
seconds at each step.
Press any key to exit ramping.
4.8 Simulating Transmitter
The function of Simulating Transmitter is only applicable to EDEKA TC-200 Multi-function
Содержание EDEKA TC-200