15
Figure6- 4 Frequency Sourcing Connections
4.5 Simulating Thermocouples
To select Simulating Thermocouples function, proceed as follows:
(1)
Press
until the lower display shows “SOURCE”.
(2)
Press
to select TC type. Continue pressing it to select the desired TC type.
(3)
Connect the test leads as shown in Figure6- 5.
(4)
Enter the desired output value by pressing the numeric keypads and press
.
CAUTION: Please pay attention to the current cold junction temperature, and if necessary, press
to
adjust the cold junction temperature manually.
4.6 Simulating RTD
To select Simulate an RTD function, proceed as follows:
(1)
Press
until the lower display shows “SOURCE”.
(2)
Press
to select RTD display. Continue pressing it to select the desired RTD
type.
(3)
Connect the test leads as shown in Figure6- 6.
(4)
Enter the desired output value by pressing the numeric keypads and press
.
CAUTION:
EDEKA TC-200 Multi-function process calibrator can only simulate 2-wire connection.
CAUTION: Clear of RTD before simulating according to ‘8.4 Clear of Resistance and RTD’.
Содержание EDEKA TC-200
Страница 1: ...EDEKA TC 200 Temperature Calibrator Huazheng Electric Manufacturing Baoding Co Ltd...
Страница 2: ......
Страница 6: ......